ArcCheck模体与胶片在鼻咽癌VMAT剂量验证中的比较研究 被引量:7

Comparison of ArcCheck and Film Verification for VMAT in the Treatment of Nasopharyngeal Carcinoma
暂未订购
导出
摘要 比较研究ArcCheck与胶片在鼻咽癌容积弧形调强(VMAT)验证中的应用,探讨ArcCheck在VMAT剂量验证中的可行性。随机选取5例鼻咽癌病人的VMAT治疗计划,分别用ArcCheck和胶片两种方法进行剂量验证,比较两种方法的验证结果。ArcCheck验证的平均通过率为96.74±2.60,胶片验证的平均通过率是95.95±1.17,利用统计分析t检验,结果表明ArcCheck和胶片验证两种结果在剂量学上没有明显差异。ArcCheck分析软件可以直接获取三维合成的剂量分布,且操作过程较胶片验证简单,可以更为便捷的进行VMAT剂量验证。 To compare the performance of ArcCheck and film verification for volumetric intensity modulated arc therapy (VMAT) in the treatment of nasopharyngeal carcinoma and to study the feasibility of ArcCheck in VMAT dosimetric verification. Five patients of nasopharyngeal carcinoma treated with VMAT were enrolled in this study. Dose verification was carried out by ArcCheck and film respectively. The result showed that there were no significant differences between ArcCheck and film verification. ArcCheck software can obtain three dimensional dose distribution directly with simple operation. It is convenient for ArcCheck to be used for VMAT dosimetric verification.
出处 《中国医疗器械杂志》 CAS 2013年第3期228-231,共4页 Chinese Journal of Medical Instrumentation
基金 温州市科技计划项目(Y20120137)
关键词 鼻咽癌 VAMT计划 ArcCheck验证 胶片验证 nasopharyngeal carcinoma, volumetric modulated arc therapy, ArcCheck verification, film verification
  • 相关文献

参考文献17

  • 1Kam MK, Teo PM, Chau RM, et al. Treatment of nasopharyngeal carcinoma with intensity-modulated radiotherapy: the Hong Kong experience[J]. Int J Radiat Oncol Biol Phys, 2004, 60(5): 1440- 1450.
  • 2Yon S, Wu Q J, Lee WR, et al. Radiotherapy treatment plans with RapidArc tbr prostate cancer involving seminal vesicles and lymph nodes[J]. Int J Radiat Oncol Biol Phys, 2010, 76: 935-942.
  • 3Feygelrnan V, Zhang G, Stevens C, et al. Evaluation of a new VMAT QA device, or the "X" and "O" array geometries[J]. J Appl Clin Med Play, 2011, 12(2): 3346.
  • 4Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc[J]. Med Phys, 2008, 35(1):310-317.
  • 5Ezzell G, Galvin J, Low D, et al treatment planning, and clinical Guidance document on delivery, mplementation of IMRT: report of the 1MRT subcommittee of the AAPM radiation therapy committee[J]. Med Phys, 2003, 30:2089-2115.
  • 6Zeidan OA, Stephenson SA, Meeks SL, et al. Characterization and use of EBT radiochromic fihn for IMRT dose verification[J]. Med Phy, 2006, 33 (11): 4064-4072.
  • 7Letourneau D, Publicover J, Moseley D J, et al. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy[J]. Med Phys, 2009, 36(5): I813-1821.
  • 8余建义,金献测,吴式,易金玲.调强放射治疗胶片验证的胶片校准[J].中国医学物理学杂志,2007,24(5):319-321. 被引量:5
  • 9Buonamici FB, Cornpagnucci A, Marrazzo L, et al.An mtercomparison between film dosimetry and diode matrix for IMRT quality assurance[J], Med Phy, 2007.34 (4): 1372-1379.
  • 10Vergote K, De Decne Y, Duthoy W, et al. Validation and application of polymer gel dosimetry for the dose verification of an intensity- modulated arc therapy (IMAT) treatment[J]. Phys Med Biol, 2004, 49(2), 287-305.

二级参考文献11

  • 1Bortfeld T,Boyer AL,Schlegel W,et al.Realization and verification of three2dimensional conformal radiotherapy with modulated fields[J].Int J Radiat Oncol Biol Phys,1994,30:899-908.
  • 2Tsai JS,Wazer DE,LingMN,et al.Dosimetric verification of the dynamic intensity modulated radiation therapy of 92 patients[J].Int J Radiat Oncol Biol Phys,1998,40:1 213-1 230.
  • 3CHAO K S, OZYIGIT G, TRAN B N, et al. Patterns of failure in patients receiving and postoperative IMRT for head- and- neck cancer [J]. Int J Radiat Oncol Biol Phys, 2003, 55 (2): 312-321.
  • 4Nathan L Childress,Lei Dong,and Isaac I.Rosen.Rapid radiographic film calibration for IMRT verification using automated MLC fields.Med.Phys,2002,29:2384-2390.
  • 5Nathan L.Childress,Mohammad Salehpour,and Lei Dong,et al.Dosimetric accuracy of Kodak EDR2 film for IMRT verifications.Med.Phys,2005,32:539-548.
  • 6C.Dancui,B.S.Proimos,and J.Roseenwald,et al.Variation of sensitometric curves of radiographic films in high energy photon beams.Med.Phys.2001,28:966-974.
  • 7XR Zhu,S Yoo,and PA Jursinic,et al.Characteristics of sensitometric curves of radiographic films.Med Phys,2003,30:912-919.
  • 8N.Dogan and G.P.Glasgow.Surface and build-up region dosimetry for obliquely incident intensity modulated radiotherapy 6MV x rays.Med.Phys,2003,30:3091-3096.
  • 9XR Zhu,PA Jursinic,and DF Grimm,et al.Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator.Med.Phys,2002,28:1687-1692.
  • 10N Dogan,LB Leybovich,and A Sethi.Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy.Phys.Med.Biol,2002,47:4121-4130.

共引文献8

同被引文献64

  • 1柏朋刚,李奇欣,张秀春,费召东,缪龙华.两种验证工具对鼻咽癌调强验证结果对比分析[J].中国医学物理学杂志,2011,28(6):2964-2966. 被引量:10
  • 2高枫.医用电子直线加速器的性能检测与质量保证[J].现代医药卫生,2006,22(6):797-798. 被引量:18
  • 3Feygelman V, Zhang G, Stevens C, et al. Evaluation of a new VMAT QA device, or the "X" and "O" array geometries[J]. J Appl Clin Med Phy, 2011, 12(2): 3346.
  • 4Mackie TR, Holmes T, Swerdloff S,et al. Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapy J]. Med Phys,1993, 20( 6): 1709-1719.
  • 5Mackie TR, Balog J, Ruchala K, et al. Tomotherapy[J]. Semin Radiat Oncol, 1999, 9(1): 108-117.
  • 6Lulent A, Mete Y, Gulbin O, et al. Total marrow irradiation with rapidare volumetric arc therapy[J]. Int J Radiat Oncol Biol Phys, 2011, 81 (2): 592-599.
  • 7Low DA, Mutic S. Abutment region dosimetry for sequential arc IMRT dlivery[J]. Phys Med Biol, 1997, 42(7): 1465-1470.
  • 8Feygelman V, Zhang G, Stevens C, et al. Evaluation of a new VMAT QA device, or the "X" and "O" array geometries[J]. Clin Med Phy, 2011, 12(2): 3346.
  • 9Gordon M Mancuso, Jonas D Fontenot, John P Gibbons, et al. Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments [J]. Med Phys, 2012, 39 (7): 4378-4383.
  • 10Nakaguchi Y, Araki F, Saiga S, et al. Comparison of dose accuracy between 2D array detectors for pre-treatment IMRT QA [J]. Med Phy, 2012, 63(4): 443-452.

引证文献7

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部