期刊文献+

过氧化氢低温等离子体灭菌器在儿科学器械灭菌中的应用 被引量:1

Application of Low Temperature Hycrogen Peroxide Plasma Sterilizer into Intrumental Disinfection in Paediatric
暂未订购
导出
摘要 目的探讨过氧化氢低温等离子在儿科器械灭菌中的灭菌效果和失败原因的观察,为正确使用等离子灭菌器提供实践指导。方法对1264次使用过氧化氢等离子低温灭菌系统的情况进行监测。结果 1244次灭菌循环运行完成,20次循环被取消。结论过氧化氢等离子灭菌质量可靠、高效、环保。 Objective To set up guidance in the right usage of the plasma sterilizer,an observation of the efficiency and failure in instrumental sterilization in paediatric with the low temperature hycrogen peroxide plasma are discussed and the causes explored.Method The process of 1244 cases of sterilization with the low temperature hycrogen peroxide plasma was put under supervision.Results The process of 20 cases was cancelled after the 1244 processes were finished.Conclusion The sterilization with the hycrogen peroxide plasma is reliable,effevtive and clean.
出处 《中国医药指南》 2013年第2期13-14,共2页 Guide of China Medicine
关键词 过氧化氢 低温 等离子 灭菌 Hycrogen peroxide Low temperature Plasma Sterilization