期刊文献+

呼吸机流量时间曲线判断气管插管后导管位置

Using ventilator flow time curve to judge catheter position after endotracheal intubation
暂未订购
导出
摘要 目的比较以呼吸机流量时间曲线和传统方法判断保留自主呼吸的患者气管插管后导管位置的快捷性和准确性。方法选择82例保留自主呼吸的急诊气管插管成年患者,随机分为流量时间曲线组和传统方法组,各41例,分别以呼吸机流量时间曲线和传统方法判断气管插管后导管位置,比较两种方法的判断时间和判断准确率。结果判断时间:流量时间曲线组3.24±1.28 s,传统方法组12.81±9.21 s,P<0.0001。准确率:流量时间曲线组100%,传统方法组97.56%,P=0.314。结论对于保留自主呼吸的气管插管患者,以呼吸机流量时间曲线判断气管插管后导管位置的快捷性优于传统方法,同时具有高度的准确性。 Objective To compare the agility and accuracy of catheter position judgement by the ventilator flow time curve and the traditional methods after endotracheal intubation. Methods 82 cases emergency intubation adult patients, whose spontaneous breath- ing were reserved, were randomly and evenly divided into the flow time curve group and the traditional method group. Then the time and accuracy of judgement were compared between the two groups. Results The judgment time of the flow time curve group was 3.24 ± 1.28 s, and the judgment time of the traditional method group was 12. 81 ±9.21 s. The difference of judgement time was significant between the two groups ( P 〈0. 0001 ). The accuracy rates was 100% in the flow time curve group and 97.56% in the traditional method group ( P = 0. 314). Conclusion The ventilator flow time curve is superior to the traditional method in the judgement of catheter position, which has the advantages of rapidity and a high degree of accuracy.
出处 《临床肺科杂志》 2013年第3期476-478,共3页 Journal of Clinical Pulmonary Medicine
关键词 呼吸机 流量时间曲线 气管内插管 位置 ventilator flow time curve endotracheal intubation position
  • 相关文献

参考文献4

  • 1Grmec S,Mally S. Prehospital determination of tracheal tube placement in severe head injury[J].Emergency Medicine Journal,2004.518-520.
  • 2P.Peeiffer,S.S.Rudolph,J.Borglum. Temporal comparison of ultrasound vs auscultation and capnography in verification of endotracheal tube placement[J].Acta Anaesthesiologica Scandinavica,2011,(10):1190-1195.
  • 3李成辉,赵薇,张金华,贾乃光.临床麻醉中几种鉴别气管导管位于气管或误插入食管的方法比较[J].中国医学科学院学报,2003,25(2):197-200. 被引量:2
  • 4G.M.Schmolzer,S.B.Hooper,K.J.Crossley. Assessment of gas flow waves for endotracheal tube placement in an ovine model of neonatal resuscitation[J].Resuscitation,2010,(06):737-741.

二级参考文献9

  • 1Bozeman WP, Hexter D, Liang HK, et al. Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation. Ann Emerg Med, 1996; 27:595-599.
  • 2Sum Ping ST, Mehta MP, Symreng T. Reliability of capnography in identifying esophageal intubation with carbonated beverage or antacid in the stomach. Anesth Analg, 1991; 73:333-337.
  • 3Keenan RL, Boyan CP. Cardiac arrest due to anesthesia:a study of incidence and causes. JAMA, 1985, 253:2373-2377.
  • 4Li J. Capnography alone is imperfect for endotracheal tube placement confirmation during emergency intubation. J Emerg Med, 2001, 20:223-229.
  • 5Grmec S. Comparison of three different methods to confirm tracheal tube placement in emergency intubation. Intensive Care Med, 2002, 28:701-704.
  • 6Sprung J, Hunter K, Barnas GM, et al. Abdominal distention is not always a sign of esophageal intubation: cardiac arrest due to “auto-PEEP”. Anesth Analg, 1994, 78: 801-8O4.
  • 7Puntervoll SA, Soreide E, Jacewicz W, et al. Rapid detection of esophageal intubation: take care when using colorimetric capnometry. Acta Anaesthesio] Scand, 2002, 46:455-457.
  • 8Asai T, Shingu K. Case report: a normal capnogram despite esophageal intubation. Can J Anaesth, 2001, 48:1025-1028.
  • 9Bigeleisen PB. An unusual presentation of end-tidal carbon dioxide after esophageal intubation. Anesth Analg, 2002,94:1534-1536.

共引文献1