心肌SPECT灌注显像的心脏轴向移动校正 被引量:2

Correction of cardiac axial movement during myocardial perfusion SPECT imaging
暂未订购
导出
摘要 目的对SPECT心肌灌注显像中心脏的轴向移动进行校正,以最大限度地降低运动伪影对诊断的影响。方法通过电影、正弦图和回旋图显示,实现对心脏移动可视的定性评估;通过帧与帧之间的互相关函数方法,找出位移角度,对投影的校正给出建议值;经可视评估后,确定投影移动校正的初值;根据移动校正的初值重建图像,通过对重建图像的投影与采集投影的互信息分析,最后计算出投影移动校正。结果心肌模型实验表明,本方法能够精确地复现模型的轴向运动,位置误差小于3mm,经校正后重建图像改善明显。结论本方法具有客观、精确、可视的优点,可精确确定和校正SPECT心肌灌注显像中心脏的轴向移动。 Objective To correct the cardiac axial movement during myocardial perfusion SPECT imaging, in order to re- duce the motion-artifacfs influence on diagnosis. Methods Visual inspection of motion of the heart was done by means of cine of the raw data, sonogram and cyclogram. By cross-correlation, the camera angles were localized at which the motion occurred and determine the suggested correction for the motion. By means of visual inspection, the correction was modified and verified. Based in the correction at this stage, the images were reconstructed. The final motion correction was then de- termined by means of maximize the cross-correlation between the projection from the reconstructed image and the measured projection. Results Myocardial phantom showed that the motion of the phantom can be revealed at an accuracy of 3 mm for various kinds of axial motions. Motion artifacts on myocardial perfusion SPECT images were effectively improved. Conclusion This approach can be used to provide non-subjective, accurate, quantitative, and visual means of axial motion correction in myocardial perfusion SPECT imaging.
出处 《中国医学影像技术》 CSCD 北大核心 2012年第12期2252-2256,共5页 Chinese Journal of Medical Imaging Technology
基金 国家重点基础研究发展计划(973计划)项目(2011CB707701) 北京市共建项目(JD100010609)
关键词 心肌灌注显像 体层摄影术 发射型计算机 单光子 伪影 校正 Myocardial perfusion imaging Tomography, emission-computed, single-photon Artifacts Correction
  • 相关文献

参考文献13

  • 1Botvinick EH, Zhu YY, O'Connel WJ, et al. A quantitative as- sessment of patient motion and its effect on myocardial perfusion SPECT imaging. J Nucl Med, 1993,34(2) :303-310.
  • 2Eisner R, Churchwell A, Noever T, et al. Quantitative analysis of the tomographic thallium-201 myocardial bullseye display: Critical role of correcting for patient motion. J Nuel Med, 1988, 29(1) :91-97.
  • 3Eisner RL, Noever T, Nowak D, et al. Use of cross-correction function to detect patient motion during SPECT imaging. J Nucl Med. 1987,28(1):97-101.
  • 4Geck[e WJ, Frank TL, Links JM, et al. Correction for patient and organ movement in SPECT: Application of exercise thallium- 201 cardiac imaging. J NuclMed, 1988,29(4) :441-450.
  • 5Cooper JA, Neumann PH, McCandless BK. Detection of patient motion during tomographic myocardial perfusion imaging. J Nucl Med, 1993,34(8):1341-1348.
  • 6Bai C, Maddahi J, Kindem J, et al. Development and evaluation of a new fully automatic motion detection and correction technique in cardiac SPECT imaging. J Nucl Cardiol, 2009, 16(4) :580-589.
  • 7Ando R, Ogawa K. Correction of patient movement with a phase --only correlation method ina SPECT study. E2012-02-20~. ht- tp://ieeexplore, ieee. org/xpls/abs _ all. jsp arnumber = 5874194 tag= 1.
  • 8Matsumoto N, Berman DS, Kavanagh PB, et al. Quantitative as sessment of motion artifacts and validation of a new motion-cor rection program for myocardial perfusion SPECT. J Nucl Med 2001,42(5) :687-694.
  • 9戚忠智,田蓉.位移对双探头SPECT心肌显像质量的影响及校正[J].四川大学学报(医学版),2011,42(3):392-396. 被引量:1
  • 10Sarkar S, Oghabian MA, Mohammadi I, et al. A linogram/si- nogram cross-correlation method for motion correction in planar and SPECT imaging. IEEE Trans Nucl Sci, 2007,54(1) :71-79.

二级参考文献20

  • 1Nigam A, Humen DP. Clinical validation of technetium99m MIBI-gated single-photon emission computed tomography (SPECT) for avoiding false positive results in patients with left bundle-branch block: comparison with stress-rest nongated SPECT[J]. Clin Cardio
  • 2Banzo I, Carril JM, Jimenez-Bonilla J, et al. Effect of attenuation correction and scatter compensation on the 99mTcMIBI myocardial perfusion spect in patients without coronary artery disease[J]. Rev Esp Med Nucl, 2002, 21(6):417-425.
  • 3Araujo W, DePuey EG, Kamran M, et al. A compari-son of TI-201 stress-reinjection-prone SPECT and Tc-99m-sestamibi gated SPECT in the differentiation of inferior wall defects from artifacts[J]. Nucl Med Commun, 2000, 21(8):719-727.
  • 4Leong LK, Kruger RL, O'Connor MK. A comparison of the uniformity requirements for SPECT image reconstruc-tion using FBP and OSEM techniques [J]. J Nucl Med Technol, 2001, 29(2): 79- 83.
  • 5Funahashi M, Shimonagata T, Mihara K, et al. Application of pixel truncation to reduce intensity artifacts in myocardial SPECT imaging with Tc-99m tetrofosmin[J]. J Nucl Cardiol, 2002, 9(6): 622-631.
  • 6Funahashi M, Shimonagata T, Mihara K, et al. 99tTcm-MIBI stress-rest myocardial perfusion scintigraphy in patients with complete left bundle branch block [J]. Cardiovasc J S Afr, 2001, 12(5): 252-256.
  • 7Nigam A, Humen DP. Prognostic value of myocardial perfusion imaging with exercise and/or dipyridamole hyperemia in patients with preexisting left bundle branch block[J]. J Nucl Med, 1998, 39(4): 579-581.
  • 8Abufadel A, Eisner RL, Schafer RW. Differences due to collimator blurring in cardiac images with use of circular and elliptic camera orbits[J]. J Nucl Cardiol, 2001, 8(4):458-465.
  • 9Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control[J]. J Nucl Med Technol, 2001, 29(1): 12-18.
  • 10Fitzgerald J, Danias PG. Effect of motion on cardiac SPECT imaging: recognition and motion correction [J]. J Nucl Cardiol, 2001, 8(6): 701-706.

共引文献12

同被引文献16

  • 1Caobelli F,Pizzocaro C,Paghera B,et al.Evaluation of patients with coronary artery disease.IQ-SPECT protocol in myocardial perfusion imaging:Preliminary resnlts[J].Nuklearmedizin, 2013,52(5):178-185.
  • 2Garcia EV,Faber TL,Esteves FP.Cardiac dedicated ultrafast SPECT cameras:new designs and clinical implications[J].JNucl Med,2011,52(2):210-217.
  • 3Horiguchi Y,Ueda T,Shiomori T,et al.Validation of short-scan- time imaging protocol for thallium-20l myocardial SPECT with a multifocal collimator.Ann Nucl Med,2014,28(8):707- 715.
  • 4Imbert L,Poussier S,Franken PR,et a/.Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT:a comprehensive analysis of phantom and human images [J] .J Nucl Med,2012,53 (12): 1897 - 1903.
  • 5Giorgetti A,Masci PG,Marras G,et al.Gated SPECT evaluation of left ventricular function using a CZT camera and a fast low- dose clinical protocol:compatison to cardiac magnetic resonance imaging[J].Eur J Nucl Med Mol Imaging,2013,40(12): 1869-1875.
  • 6Cochet H,Bullier E,Gerbaud E,et al.Absolute quantification of left ventricular global and regional function at nuclear MPI using ultrafast CZT SPECT:initial validation versus cardiac MR[J]. Nucl Med,2013,54(4):556-563.
  • 7Duvall WL,Croft LB,Ginsberg ES,et aLP, educed isotope dose and imaging time with a high efficiency CZT SPECT camera[J].JNucl Cardiol,2011,18(5):847 -857.
  • 8张洁,马丽,李顺华,陈正福.SPECT仪器质量控制和质量保证的初步探讨[J].中国临床医学影像杂志,2009,20(1):67-68. 被引量:29
  • 9王渊恺,张光明,韩芳,刘从进,刘兴党.不同采集矩阵与函数重建对SPECT影像质量影响[J].中国医学计算机成像杂志,2011,17(2):164-167. 被引量:9
  • 10任志刚,李小华,刘枫,刘淑华.SPECT的数据采集条件对其断层分辨率的影响实验探讨[J].中国医学物理学杂志,2000,17(1):4-5. 被引量:8

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部