期刊文献+

基于误差与抗干扰优化的中频强磁场测量仪设计 被引量:2

Design of a Strong Magnetic Field Measurement Based on Error and Interference Optimization
在线阅读 下载PDF
导出
摘要 通过对中频强磁场空间点测量工况分析和信号处理电路误差计算,找出了产生误差的主要因素.为提高测量仪的精度、实时性和抗干扰能力,采用软硬件相结合优化方法,在信号调理电路中,合理配置前后级的放大倍数和相关电阻以及采用高精度运放,可以大大减少小信号放大引起的误差.采用两种硬件措施和独特平均值滤波方法可以有效抑制信号干扰.通过最小二乘法优化数据拟合参数,进一步减少了整流滤波和程控放大电路引起的误差,整体提高测量仪器性能.对优化后测量仪实验测试结果表明:测量误差小于±1.5%,一致性误差小于1%,测量的实时性也得到提高,取得了较好效果. By analysing the operating condition of space point measurement of IF strong magnetic field and computing the error of signal processing circuits, it finds out the major factors leading to errors. In order to improve the accuracy of the measuring instrument, real time property and anti-interference ability, an optimization method of combining hardware and software is presented. In the signal regulating circuit, the errors were significantly reduced by reasonably configuring the amplification factors of pre and post stage, relative resistor and using high-precision operational amplifier. Interference was suppressed effectively by employing two hardware and uniquely mean filtering methods. Via the use of the least square method to optimize data fitting parameters, errors caused by rectifier, filter and programmable am- plifier circuit were reduced further, which improved the whole performance. The experimental results obtained by testing the optimized measuring instrument, show that measurement error is less than ± 1.5%, consistency error is less than 1%, and real time property of measurement is also improved, which demonstrates the excellent effects.
出处 《广东工业大学学报》 CAS 2012年第4期27-32,共6页 Journal of Guangdong University of Technology
基金 国家自然科学基金资助项目(10475020)
关键词 中频强磁场 误差分析 优化设计 平均值滤波 放大倍数 intermediate frequency strong magnetic field error analysis optimizing design average filter magnification
  • 相关文献

参考文献14

二级参考文献15

共引文献29

同被引文献16

  • 1汤仕平,蒋全兴,周忠元,景莘慧.宽带GTEM小室场强校准装置和不确定度评定[J].电子测量与仪器学报,2004,18(Z1):543-548.
  • 2谢鸣,王维龙,黄攀,马晓庆.10MHz~1GHz电场探头校准系统[J].安全与电磁兼容,2007(5):23-27. 被引量:6
  • 3EMC WG-Working Group for EMC. IEEE Std 1309?-2005IEEE Standard for Calibration of Electromagnetic Field Sen-sors and Probes, Excluding Antennas, from 9kHz to 40GHz[S].IEEE,2005.
  • 4MYRON L C. Generation of standard EM fields using TEMtransmission cells [ J ]. IEEE Transactions on Electromagnet-ic Compatibility, 1974,16(4) : 189-195.
  • 5International Electrotechnical Commission. IEC 6100-4-20Electromagnetic Compatibility ( E-MC ) -part4 : Testing andmeasurement techniques. Section 20 : Emission and Immu-nity Testing in Transverse Electromagnetic ( TEM )Waveguides[ S]. Geneva,Switzerland,2003.
  • 6CRAWFORD M L, WORKMAN J L,THOMAS C L. Ex-panding the bandwidth of TEM cells for EMC measurem-ents [J]. IEEE Transactions on EMC, 1978,20(3) :368-375.
  • 7全国法制计量管理计量技术委员会.JJF1059.1-2012测量不确定度的评定与表示[S].北京:中国标准出版社,2012.
  • 8李梦琉.Cu-Ni-Zn合金常量成分测定及误差分析[D].广州:广东工业大学材料与能源学院,2006.
  • 9胡玉生,朱剑英,蒋全兴.金属线列横电磁波小室的场分布分析[J].电波科学学报,2008,23(1):6-10. 被引量:2
  • 10胡玉生,朱剑英.横电磁波室试验场强评估的影响因素分析[J].计量学报,2008,29(4):365-368. 被引量:2