辐射探测器定量涂硼技术 被引量:1

Technics of Quantitative Boron-Coating for Radiation Detectors
在线阅读 下载PDF
导出
摘要 报道了2种廉价实用的辐射探测器定量涂硼技术——浸没涂硼和手工刷涂,并给出了详细的涂抹细节。用不同厚度聚乙烯管包裹该方法研制的涂硼正比计数管,经标准Am-Be中子源辐照后测得了其对中子计数率的响应。浸没涂硼脂硼比λ最小值为5.0,手工刷涂λ最佳值为0.2;最佳方法是先在待涂面上滴混合液,再撒上硼粉刷涂。浸没涂硼对待涂面无限制,且硼膜均匀、牢固性都较好,但因λ较大而降低探测效率,且浸涂次数需实验和经验确定,还会造成浪费。手工刷涂克服了浸没涂硼的缺点,但对待涂表面形状有要求,且其硼膜的均匀、牢固性总体不如浸没涂硼。两种涂抹工艺操作简单、成本低廉,且能达到要求,可以在国内相关单位推广使用。 Two economic and practical technics of quantitative boron coating for radiation detectors which named dip coated boron deposition(DCBD)and hand painted boron deposition(HPBD)were presented,and the concrete steps are detailed.After been wrapped by tubes with different thickness and exposed by stand ard Am Be neutron source,response of neutron count rate of a boron lined proportional counter(BLPC)in which the boron film are painted by this way are measured.Minimum of resin boron ratio(k)in DCBD is 5.0,while the best value in HPBD is 0.2.The best method is that mixture be dropped on the coated surface first ly,then print boron powder by hand.DCBD have merits of uniform and firm boron film,and unlimited to coa ted surface,but it will lower the detection efficiency because of a big value of Test numbers of DCBD should be determined by experiments and experience,which would waste materials inevitably.HPBD overcomes the demerits of DCBD,but there are some restrictions to coated surface,as a whole,the uniformity and robust ness of boron film are not as good as DCBD.Both DCBD and HPBD show simply operation and low cost,which can meet the demands of general gas detectors and can be popularized in domestic research institutions.
作者 陈国云 魏志勇 黄福成 辛勇 朱立 CHEN Guo–yun;WIE Zhi–yong;HUANG Fu–cheng;XIN Yong;ZHU Li(Department of Physics,Nanchang University,Nanchang,Jiangxi,330031;Space Environment Center,Nanjing University of Aeronautics&Astronautics,Nanjing,Jiangsu 210016)
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2012年第9期1066-1069,1084,共5页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金(41064002)
关键词 浸没涂硼 手工刷涂 脂硼比 中子 dip coated boron deposition(DCBD) hand painted boron deposition(HPBD) resin-boron rati-o neutrons
  • 相关文献

参考文献4

  • 1丁大钊,叶春堂,赵志祥,等.中子物理学(上册)[M].北京:原子能出版社,2006:92.
  • 2陈国云,魏志勇,辛勇,方美华,黄三玻,黄国庆.涂硼电离室中子探测效率和灵敏度[J].原子能科学技术,2011,45(2):244-249. 被引量:18
  • 3P. M. Dighe, D. N. Prasad. K. R. Prasad, et al. Boron - lined proportional counters with improved neu- tron sensitivity [ J ]. Nuclear Instruments and Methods A, 2003,496( 1 ) : 154 - 161.
  • 4杨道广,陆双桐,宣肇祥.DL123型六节中子电离室研制[J].核电子学与探测技术,2008,28(3):456-461. 被引量:6

二级参考文献14

  • 1廖俊辉,谢一冈,陈元柏,郭萍,郭兰英,欧阳群,凌球,陆双桐.用于热中子探测的正比管性能研究[J].核电子学与探测技术,2007,27(2):367-371. 被引量:14
  • 2丁大钊 叶春堂 赵志祥 等.中子物理学[M].北京:原子能出版社,2001.54-59.
  • 3吴志华,赵国华等,原子核物理实验方法[M].北京:原子能出版社.1997.8-12.
  • 4丁大钊,叶春堂,赵志祥,等.中子物理学(上)[M].北京:原子能出版社,2005:92.
  • 5GUNTER P. Neutron sensitivity of a polyethy lene-walled electret ionization chamber[J]. Nu clear Instruments and Methods in Physics Re search A, 1985, 234(1): 183-190.
  • 6ENDO S, HOSHI M, SUGA S, et al. Determination of the relative neutron sensitivity of a C CO2 ionization ehamber[J]. Physics in Medicine and Biology, 1996, 41(6) : 1 037-1 043.
  • 7NANJO H, BANDO T, HASUKO K, et al. Neutron sensitivity of thin dap chambers[J]. Nu clear Instruments and Methods in Physics Re search A, 2005, 543: 441-453.
  • 8BOLOZDYNYA A, BOLOTNIKOV A, RICHARDS J, et al. Detection of thermal neutrons in cylindrical ionization chamber filled with highpressure Xe+^3 He gas mixture[J]. Nuclear Instruments and Methods in Physics Research A, 2004, 522: 595-597.
  • 9GLENN F K. Radiation detection and measurement [M]. Second Edition. Quebec, Canada: Quebecor Printing, 2000: 521.
  • 10DIGHE P M, PRASAD D N, PRASAD K R, et al. Boron-lined proportional counters with im- proved neutron sensitivity [J]. Nuclear Instruments and Methods in Physics Research A, 2003, 496: 154-161.

共引文献19

同被引文献9

  • 1丁大钊,叶春堂,赵志祥,等.中子物理学:上册[M].北京:原子能出版社,2001:56.
  • 2BADHWAR G D. Measurements on board the space shuttle[J]. Neutron, 2001, 33: 235-241.
  • 3KOSHIISHI H, MATSUMOTO H, CHISHIKI A, et al. Evaluation of the neutron radiation en- vironment inside the International Space Station based on the Bonner hail neutron detector experi- ment[JT. Radiation Measurements, 2007, 42: 1 510-1 520.
  • 4KOSHIISHI H, GOKA T, KOGA K, et al. Real-time measurement of low-energy range neu- tron spectra on board the space shuttle STS-89 (S/MM-8)[J]. Radiation Measurements, 2001, 33: 321-333.
  • 5GARNY S, MARES V, RUHM W. Response functions of a Bonner sphere spectrometer calcu- lated with Geant4[J]. Nuclear Instruments and Methods in Physics Research A, 2009, 604: 612-617.
  • 6BRAMBLETT R L, EWING R I, BONNER TW. A new type of neutron spectrometer[J]. Nucle- ar Instruments and Methods, 1960, 9(1): 1-12.
  • 7宋仔峰.基于GEM的热中子位置灵敏探测器的研究[D].合肥:中国科学技术大学,2010.
  • 8Geant4 user's guide for tookit developers [R/ OL]. 2007. http://geant4, web. cem. ch/geant4/ UserDocumentation/UsersGuides/ForToolkitDe- veloper/html/index, html.
  • 9王鑫,谷德山,刘林茂,程道文,魏晓云,贺平.D-T中子发生器慢化体结构的蒙特卡罗方法研究[J].核电子学与探测技术,2008,28(5):958-960. 被引量:4

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部