摘要
目的探讨Sysmex XS-1000i血液分析仪血小板相关参数缺失的原因及解决方法。方法将123例Sysmex XS-1000i血液分析仪血小板相关参数缺失的标本进行涂片、瑞氏染色、显微镜观察红细胞和血小板异常形态。结果在123例Sysmex XS-1000i血液分析仪血小板相关参数缺失标本中,镜检发现有不同程度的异常形态红细胞或(和)异常形态血小板增多。结论遇到Sysmex XS-1000i血液分析仪血小板相关参数缺失时,应采用瑞氏染色显微镜检查方法作为补充,对检验结果做详细描述。有条件的实验室可以用参考方法检测血小板。
Objective To investigate the cause and solution of no platelet-associated parameters displayed by the Sysmex XS-1000i hematology analyzer.Methods Swear,Wright Stain,then microscopic examination the red blood cells and platelets abnormal morphology in 123 cases of no platelet-associated parameters were analyzed by the Sysmex XS-1000i hematology analyzer.Results In 123 cases of no platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer,microscopic examination showed varying degrees of abnormal morphology of red blood cells and(or) abnormal morphology thrombocytosis.Conclusion Encountered the lack of platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer must use the method of Wright stain and microscopic examination as a supplement,which can provide a detailed description of the test results.Platelet can be detected using reference method in laboratory conditions.
出处
《检验医学与临床》
CAS
2012年第22期2830-2831,2833,共3页
Laboratory Medicine and Clinic
关键词
血液分析仪
血小板相关参数缺失
红细胞
血小板形态
镜检
血小板计数
hematology analyzer
platelet-associated parameters deficiency
abnormal morphology of red blood cells and platelets
microscopic examination
reference method of platelet count