ORTHO Innov全自动分析仪在血型鉴定中的性能评价 被引量:1

Performance Evaluation of Bloodgrouping with Automatic ORTHO Innova Analyzer
暂未订购
导出
摘要 目的对ORTHO Innova全自动血型及配血分析系统在ABO血型鉴定中的性能进行评价。方法对12 463例合格标本进行仪器法和试管法ABO血型鉴定以评价准确度;选取AB型、O型各10例,已确认为B3型1例、Ax型2例标本,评价精密度;选取合格和不合格各类标本各3例以评价交叉污染;并探讨不合格标本的影响情况。结果仪器法准确度与试管法相比差异无统计学意义(P>0.05);仪器法对正常血型标本重复检测完全符合;对亚型检测时偶有0.5+的误差;有较大漂浮凝块标本的携带污染率为100.00%,其他标本携带污染率为0;不合格标本对检测结果有不同程度影响。结论该仪器在正常ABO血型鉴定中准确度和精密度高,无携带污染,但使用标本必须合格。 Objective To evaluate automatic ORTHO Innova analyzer for identification of blood groups of ABO. Methods ABO blood groups were identified with blood group analyzer and tube method for accuracy evaluation in 12 463 qualified samples. Each 10 cases of AB and O group, 1 case of B3 group, 2 cases of Ax group were tested for precision e- valuation. Each 3 qualified and unqualified samples were tested for carry-over rates. Simultaneously, to explore the effect of unqualified samples on results. Results Compared with tube method,the accuracy of blood group analyzer was not dif- ferent (P^0. 05). In bloodgrouping with normal ABO group samples,the results of blood group analyzer was entirely co- incidental in the repeating test. Occasionally wrong with 0. 5+ existed in subgroup test. The carry-over rate was 100. 00% in samples with bigger floating blood clots,but 0 in other samples. There were different influences on bloodgrouping in ab- normal samples. Conclusion The accuracy and precision of blood group analyzer are high for normal ABO bloodgrouping, no carry over,but the sample must be qualified.
作者 徐朴 余华
出处 《临床输血与检验》 CAS 2012年第4期314-316,共3页 Journal of Clinical Transfusion and Laboratory Medicine
关键词 全自动血型分析仪 准确度 精密度 交叉污染率 Automatic blood group analyzer Accuracy Precision Carry-over rate
  • 相关文献

参考文献9

二级参考文献15

共引文献17

同被引文献4

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部