苏州市某医院介入操作场所辐射水平调查

Survey of radiation levels in interventional operation place of a hospital in Suzhou City
原文传递
导出
摘要 目的了解介入手术操作现场的辐射水平,并估算介入操作人员的年剂量水平。方法采用BH3103A便携式X-γ剂量率仪检测介入操作现场辐射剂量率。结果防护屏内外侧的吸收剂量率分别为(38.59±7.48)和(10.52±5.86)μGy/h,医生防护服内外侧的吸收剂量率分别为(3.06±1.17)和(75.35±2.63)μGy/h,初步估计医生年有效剂量约为0.334 mSv/a。防护屏的防护效果为72.7%,防护服的防护效果为95.9%。结论医生年有效剂量小于国家规定限值。 To understand the radiation level of interventional operation place, and estimate its annual dose level. [ Methods] BH3103A Portable X-γ dose rate monitor was used to detect the radiation dose rate of interventional operation place. [Results] The absorbed dose rate of shield inside and outside was (38. 59 ±7. 48) μGy/h and (10. 52 ±5. 86} μGy/h, respectively; that of medical protection clothing inside and outside was ( 3. 06 ± 1. 17 ) μGy/h and (75. 35 ± 2. 63 ) μGy/h, respectively. The pre-estimated effective dose was about 0. 334 mSv/a annually. The protective effect of shield and protective clothing was 72. 7% and 95.9%, respectively. [ Conclusion] The annual effective dose of doctor is lower than the national limits.
出处 《职业与健康》 CAS 2012年第17期2093-2094,共2页 Occupation and Health
关键词 介入诊治 辐射防护 医学照射 累积剂量 Interventional diagnosis and treatment Radiation protection Medical irradiation Cumulative dose
  • 相关文献

参考文献5

二级参考文献30

  • 1孟冰,闫琳,由京周.介入放射工作者受照剂量调查及防护措施应用[J].职业卫生与应急救援,2004,22(3):133-133. 被引量:5
  • 2陶永军,陈本宝,徐孝波,傅立军,王红飞.介入放射手术时被检者和操作者受照射剂量调查[J].中华放射医学与防护杂志,2005,25(1):76-77. 被引量:8
  • 3UNSCEAR.1993年报告 电离辐射源与效应[R].北京:原子能出版社,1995..
  • 4.GBZ 130—2002.医用X射线诊断卫生防护标准[S].,2002..
  • 5Padovani R, Rodella CA. Staff Dosimetry in Interventional Cardiology [ J ]. Radiation Dosimetry,2001,94 ( 1 - 2 ) : 99 - 103.
  • 6Kicken PJH, Kemerink GJ, Van Engelshoven JMA. Dosimetry of occupationally exposed persons in diagnostic and interventional arteriography. Part I: Assessment of Entrance Doses [J]. Radiation Dosimetry , 1999,82(2) :99 - 103.
  • 7Kicken PJH, Kemerink GJ, Schultz FW, et al. Dosimetry of occupationally exposed persons in diagnostic and interventional arteriography. Part Ⅱ: Assessment of Entrance Doses [ J ]. Radiation Dosimetry, 1999,82 ( 2 ) : 105 - 114.
  • 8Casey M, Malone DE. The value of digital personal dosemeters in angiography/interventional radiology: Preliminary Experience[J]. Radiation Dosimetry, 2001,94 ( 1 - 2 ) : 93 - 94.
  • 9Kottou S, Neofotistou V, Tsapaki V. Personnel dosed in haemodynamic Units in Greece [ J ]. Radiation Dosimetry ,2001, 94( 1 - 2) : 121 - 124.
  • 10Janssen RJ JN, Hadders R J, Henkelman RH, et al. Exposure to operating staff during cardiac catheterization measured by thermoluminescence Dosimetry[J]. Radiation Dosimetry, 1992,43(1) :175 - 177.

共引文献164

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部