期刊文献+

氢钟守时应用 被引量:6

Study on the usage of hydrogen clock in time keeping
在线阅读 下载PDF
导出
摘要 通过对氢钟时差测量数据的二次差分,获得频率漂移的准确量值后对氢钟时差测量数据进行修正,经过频率漂移修正后的氢钟数据在保持其良好短稳特性的基础上长稳性能得到了较大提升,达到或超过性能中上的高性能管的商品铯钟5071A的常稳指标,这种处理方法能提高守时氢钟对时间实验室所保持的时间尺度的贡献。 Based on the second difference of the measured time-difference for the hydrogen clock,the frequency drift of hydrogen clock can be estimated and corrected.The corrected data for the hydrogen clock not only keep the good short-term stability but also show much better long-term stability which achieves or surpasses that of HP5071A caesium clock,a caesium clock with middle-upper level.Thus the contribution of hydrogen clock to the time scale kept in time laboratories can be enhanced if the method discussed in this paper is adopted.
作者 屈俐俐
出处 《时间频率学报》 CSCD 2012年第2期82-87,共6页 Journal of Time and Frequency
基金 国家自然科学基金资助项目(11103025)
关键词 氢钟 频率漂移 守时 hydrogen clock frequency drift time-keeping
  • 相关文献

参考文献4

  • 1Agilent technologies. Agilent 5071A Primary Frequency Standard Operation and Programming Manual[K]. 2009.
  • 2Symmetricom Sigma Tau Standards Group. The Model MHM-2010 Atomic Hydrogen Maser Frequency Standard Instruction and Operations Manual[K]. 2007.
  • 3彭科,张燕军,张为群,张一平,蔡勇,翟造成.SOHM-4型氢原子钟的设计改进与初步性能[J].时间频率学报,2004,27(1):41-47. 被引量:8
  • 4OWJ Riley Hamilton Technical Services. Stable32 Frequency Stability Analysis user manual[K]. 2007.

二级参考文献4

共引文献7

同被引文献27

引证文献6

二级引证文献23