全自动清洗消毒器清洗效果的研究 被引量:8

Verification of cleaning effect of automatic washer-disinfector
原文传递
导出
摘要 目的探讨影响清洗质量的相关因素及清洗效果验证的方法。方法利用目测、ATP检测、清洗监测卡监测、温度验证及过程参数确认的方法,对清洗机程序的清洗效果进行验证。结果清洗剂浓度、清洗温度和时间对最终清洗效果产生影响;不同品牌清洗机使用同一清洗剂,达到合格清洗效果要求的浓度、温度、时间也可能不同;同一台清洗机使用不同清洗剂,达到合格清洗效果要求的浓度、温度、时间也可能不同;同一型号使用年限相同的老机器,使用同一清洗剂、同一程序,到达合格清洗效果要求的浓度、温度、时间也可能不同;新设备出厂验证设置与消毒供应中心实际情况不一致,清洗效果测试不合格。结论当清洗消毒器安装、更新、大修、更换清洗剂、清洗程序选择发生改变时需要对清洗效果进行检测;应加强清洗消毒机的日常管理和监测。 OBJECTIVE To explore the impact factors for the cleaning efficacy and the measure to verify the cleaning effect of the automatic washer-disinfectors. METHODS The cleaning effects of the procedures of the automatic washer-disinfector were verified by visual inspection, ATP determination, cleaning monitoring card, temperature verification, and process parameters determination. RESULTS The detergent concentration, cleaning temperature, and cleaning duration could impact the final cleaning effect. The different washer-disinfector with the same detergents varied in the concentration, temperature, and the duration to meet the qualified cleaning effect. The same wash-disinfectors with different detergents may varied in the concentration, temperature, and the duration to meet the qualified cleaning effect. The washer-disinfectors, with the same model, same procedure ,and same detergent, may vary in the concentration, temperature, and the duration to meet the qualified cleaning effect. The factory output verification settings of the newly devices were not exactly same as the CSSD, the cleaning effects were tested unqualified. CONCLUSION It is necessary to monitor the cleaning effect of the washer- disinfector upon its installation, major repairs, and the change of detergent or the cleaning procedure. The daily management and monitoring of the washer-disinfector should be strengthened.
作者 张青 高宏
出处 《中华医院感染学杂志》 CAS CSCD 北大核心 2012年第16期3592-3594,共3页 Chinese Journal of Nosocomiology
关键词 全自动清洗消毒器 清洗效果 验证 Automatic washer-disinfector Cleaning efficacy Verification
  • 相关文献

参考文献3

二级参考文献22

  • 1郑柏湘,龚科.消毒供应室风险管理评估及防范措施[J].中华医院感染学杂志,2007,17(7):854-856. 被引量:30
  • 2中华人民共和国卫生部.中华人民共和国卫生行业标准[S].2009:1-6.
  • 3Murphy D, Whiting J. Dispelling the Myths: the true cost of health-associated infections[J]. An APIC briefing,2007.
  • 4R. Douglas Scott II. The direct medical costs of healthcare-associated infections in U.S. hospitals and the benefits of prevention[S]. Centers for dieases control and prevention,2006.
  • 5Doyle JE, Ernst RR. Resistance of Bacillus subtilis var. niger Spores Occluded in Water-insoluble Crystals to Three Sterilization Agents[J]. Appl microbiol, 1967,15(4) : 726-730.
  • 6Alfa MJ. New low temperature sterilization technologies:mi crobicidal activity and clinical efficacy In:Rutala WA, ed[J]. Disinfection,sterilization, and antisepsis in healthcare, 1998, 67-78.
  • 7Charles L. Mullican, Robert K. Hoffman. Dry Heat or Gase ous Chemical Resistance of Bacillus subtilis var. niger Spores Included Within Water-soluble Crystals[J]. Applied Microbiology, 1968,16(8): 1110-1113.
  • 8Diab-Elschahawi M, Balcky A, Bachhofner N, et al. Challeng ing the Sterrad 100NX sterilizer with different carrier materials and wrappings under experimental " clean" and " dirty conditions[J]. Am J Infect Control,2010,38(10) :806-810.
  • 9AAMI ST79 2006. Comprehensive guide to steam sterilization and sterility assurance in health care facilities [S]. AAMI, 2006.
  • 10Rutala WA,Weber DJ. Guideline for disinfection and sterilization in healthcare facilities[S]. Center for disease control and prevention, 2008.

共引文献33

同被引文献46

引证文献8

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部