不同管电压64层螺旋CT对儿童CTA图像质量和辐射剂量的影响 被引量:14

Impact of different tube voltage protocols on image quality and radiation dosage for pediatric 64-slice cardiovascular CT angiography
暂未订购
导出
摘要 目的用64层螺旋CT观察不同管电压(80kVp及100kVp)对儿童CTA图像质量(IQ)和辐射剂量(ED)的影响。方法回顾性分析52例先天性心脏病患儿心脏CTA检查资料,根据扫描管电压分为80kVp组(n=30)及100kVp组(n=22),均采用自动管电流调制技术。采用bolus tracking技术自动注射对比剂,记录增强后心腔和大血管的密度,测量背景噪声,计算SNR、对比噪声比(CNR)及有效剂量(ED)。由2名医师采用单盲法评估图像质量。结果 80kVp和100kVp两组心腔大血管平均密度、SNR及CNR差异均无统计学意义(P均>0.05);IQ评分分别为4.17±0.70、4.23±0.53(P=0.834);80kVp组平均ED为(0.458±0.085)mSv,低于100kVp组的(1.074±0.290)mSv(P<0.001)。结论儿童心脏CTA检查中,与100kVp相比,80kVp管电压可明显降低ED而不影响IQ。 Objective To evaluate the impact of 80 and 100 kVp tube voltage on image quality(IQ) and radiation dosage in pediatric cardiovascular CT angiography(CTA) with 64-slice CT.Methods CTA was performed in 52 children with suspected congenital heart disease.Children were randomly assigned to 80 kVp(n=30) or 100 kVp(n=22) group.Other scanning parameters were kept constant.Contrast medium was injected automatically with bolus tracking.Enhancement of heart chamber and great vessels and SNR,contrast-to-noise ratio(CNR) and effective dose(ED) were calculated.Subjective vessel contrast was assessed by two radiologists in consensus.Results There was no statistical difference of mean vessel attenuation,SNR and CNR between the two groups(all P0.05),nor of subjective IQ(4.17±0.70 vs 4.23±0.53,P=0.834).Mean ED for 80 kVp group([0.458±0.085]mSv) was significantly lower than that for 100 kVp group([1.074±0.290]mSv,P0.001).Conclusion Reduction of tube voltage from 100 to 80 kVp can result in significant decrease of ED in pediatric cardiovascular CTA without deteriorating image quality.
出处 《中国医学影像技术》 CSCD 北大核心 2012年第6期1213-1217,共5页 Chinese Journal of Medical Imaging Technology
关键词 儿童 心脏缺损 先天性 体层摄影术 X线计算机 血管造影术 图像质量 辐射剂量 Child Heart defects congenital Tomography X-ray computed Angiography Image quality Radiation dosage
  • 相关文献

参考文献15

二级参考文献69

  • 1刘建新,孙红霞,唐光健,蒋学祥.多层螺旋CT低剂量对比剂肺动脉成像[J].中国医学影像技术,2006,22(7):1012-1014. 被引量:23
  • 2杨珂,张兴太,李福生,陈英民,李连波,于夕荣,柳怡,李炜,张显鹏.儿童CT扫描体表剂量水平与分析[J].中国辐射卫生,2006,15(4):433-433. 被引量:6
  • 3Colang JE, Killion JB, Vano E. Patient dose from CT: a literature review. Radiol Technol, 2007,79 : 17-26.
  • 4Brenner, DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med, 2007, 357: 2277 -2284.
  • 5Brody AS, Frush DP, Huda W, et al. Radiation risk to children from computed tomography. Pediatrics ,2007,120:677-682.
  • 6Kalra MK, Rizzo S,Maher MM,et all Chest CT performed with Z-axis modulation : scanning protocol and radiation dose. Radiology, 2005,237 : 303 -308.
  • 7Namasivayam S, Kalra MK, Pottala KM, et al. Optimization of Z-axis automatic exposure control for multideteetor row CT evaluation of neck and comparison with fixed tube current technique for image quality and radiation dose. AJNR,2006,27: 2221-2225.
  • 8Kalra MK, Maher MM, Kamath RS, et al. Sixteen-detector row CT of abdomen and pelvis: study for optimization of Z-axis modulation technique performed in 153 patients. Radiology,2004, 233 : 241-249.
  • 9Kalra MK, Maher MM, D'Souza RV,et al. Detection of urinary tract stones at low-radiation-dose CT with Z-axis automatic tube current modulation: phantom and clinical studies. Radiology, 2005,235:523-529.
  • 10Naidich DP, Marshall CH, Gribbin C, et al. Low-dose CT of the lungs : preliminary observations. Radiology, 1990,175:729-731.

共引文献227

同被引文献137

引证文献14

二级引证文献52

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部