期刊文献+

新鲜血清重复测定差值法在游离甲状腺激素质量控制中的应用

Application of Repetitive Measurement Difference Value with Patient Fresh Serum for the Quality Control of Thyroid Hormones
暂未订购
导出
摘要 目的探讨建立一种以病人新鲜血清为质控物的全自动电化学发光免疫分析仪的质控方法。方法以Roche公司生产的Cobase601全自动电化学发光分析仪为实验仪器,以血清Fr3.FT4和TStt作观察指标。每天上午在质控物随标本测定后,选取两份结果合适的新鲜病人标本,第二天重复测定1次(间隔24h),连续检测30天.计算FT3,FT4和TsH前后2次检测结果的差值。观察其最大偏差是否满足〈1/2澳大利亚室问质评标准的要求。结果新鲜血清三项指标30天重复测定结果的平均偏差及最大偏差均〈1/2澳大利亚室间质评标准,能满足临床要求。结论病人新鲜血清标本重复测定差值法经济、方便、可行,可作为FT3,FT4和TSH室内质拉品实际应用中的补充,利于提高检验报告的质量。 Objective To set up a repetitive measurement difference value quality control method with patient fresh serum for thyroid hormones. Methods All tests were performed on Roche Cobas e601 patient fresh serum samples were selected for 24 hours later detection after quality control passed which were checked among the patient samples every morning lasted for 30 days. Counted the deviation of repetitive measurement difference value of FT3,FT4 and TSH, observed the maximum de- viation if suffice for less than a half of the request of Australia EQA standard. Results The average deviation and the maxi- mum deviation were less than a half of the request of Australia EQA standard for 30 days. It could suffice for clinic need. Conclusion Difference value method based on patient fresh serum can be used as an appending method for FT3, FT4 and TSH routine matching control material, which is a financial, convenient, feasible method, benefit to increase of quality for test result.
出处 《现代检验医学杂志》 CAS 2012年第2期150-151,154,共3页 Journal of Modern Laboratory Medicine
关键词 电化学发光免疫分析 质量控制 新鲜血清 差值法 ECLIA quality control fresh serum difference value method
  • 相关文献

参考文献6

二级参考文献10

  • 1曾宇娟,邹燕,冯树星,刘光金.冻干质控血清复溶后冰格保存的稳定性探讨[J].宜春学院学报,2006,28(2):90-91. 被引量:4
  • 2郭书云,朱爱萍.血清及两种质控品冷冻保存对生化检测结果的稳定性研究[J].临床医药实践,2006,15(10):776-777. 被引量:6
  • 3王治国 王清涛.质控方法的设计[A].见:杨振华.临床实验室质量管理[C].北京:人民卫生出版社,2003.71-78.
  • 4王治国 王清涛.质控图和质控规则[A].见:杨振华.临床实验室质量管理[C].北京:人民卫生出版社,2003.61-64.
  • 5王治国.误差和检测系统的性能要求[A].见:杨振华.临床实验室质量管理[C].北京:人民卫生出版社,2003.49-51.
  • 6魏吴 丛玉隆.医学实验室质量管理与认可指南[M].北京:中国计量出版社,2004.72-75.
  • 7International Organization for Standardization. General requirements for the competence of testing and calibration laboratories. ISO/IEC 17025,International Organization for Standardization, Geneva, 1999.
  • 8The National Conmmittee for Clinical Laboratory Standards. Method Comparison and Bias Estimation Using Patient Sampale;Approved Guideline,NCCLS, EP9-A, 1995.
  • 9邴圣民译.放射免疫测定质量控制[M].上海:上海科学技术文献出版社,1985:110.
  • 10薛晓明.质控血清的不同保存方法对测定结果的影响[J].中国血液流变学杂志,2002,12(4):360-361. 被引量:4

共引文献80