期刊文献+

结合硬斑块特征的心血管内超声图像中-外膜边缘检测 被引量:7

Hard Plaque Characteristics Based Media-Adventitia Border Detection in Intravascular Ultrasound Images
暂未订购
导出
摘要 针对心血管内超声(IVUS)图像中-外膜(media-adventitia)边缘检测中,硬斑块和声影等造成中-外膜边缘难以准确检测的问题,提出一种结合硬斑块特征的中-外膜边缘检测方法。首先,采用k均值聚类分割IVUS图像,根据图像中不同类型斑块的成像特点检测硬斑块位置;然后,计算IVUS图像方向梯度,结合斑块位置和图像灰度,获得代价矩阵;最后,利用启发式图搜索方法,实现IVUS图像的中-外膜边缘检测。对临床图像的实验结果表明,本方法能克服图像中声影和斑块干扰等问题,使检测正确率达到95.57%,提高了IVUS图像中-外膜边缘检测的准确性。 This paper presents a new approach for the detection of media-adventitia border in intravascular ultrasound(IVUS) images,to overcome problems of hard plaque and acoustic shadow which make media-adventitia difficult to detect.The approach started with image classification using k-means algorithm,such that the region of hard plaques was delineated from the IVUS image.In a next step,a cost matrix was created by combining image gradient,hard plaque location and image intensity.A heuristic graph-searching was then applied to find the media-adventitia border from the cost matrix.Experiment results showed that the approach improved the accuracy of media-adventitia border detection by overcoming problems caused by acoustic shadow and hard plaque inhomogeneity,and the correct rate reached 95.57%.
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 2012年第1期25-31,共7页 Chinese Journal of Biomedical Engineering
基金 国家自然科学青年基金(81000642) 国家自然科学基金(60672115)
关键词 心血管内超声 中-外膜边缘 硬斑块 启发式图搜索 intravascular ultrasound media-adventitia border hard plaque heuristic graph-searching
  • 相关文献

参考文献11

  • 1Mintz GS,Nissen SE,Anderson WD,et al.American college of cardiology clinical expert consensus document on standards for acquisition, measurement and reporting of Intravascular Ultrasound studies (IVUS)[J].Journal of the American College of Cardiology,2001,37(5):1478-1492.
  • 2孙丰荣,李艳玲,曲怀敬,刘泽,张梅.基于活动轮廓模型和边缘对比度特征量的血管内超声图像边缘提取[J].中国生物医学工程学报,2006,25(4):385-389. 被引量:4
  • 3孙丰荣,刘泽,李艳玲,曲怀敬,姚桂华,张运.一种改进的自适应形变模型及其血管内超声图像边缘提取应用[J].中国生物医学工程学报,2008,27(2):276-281. 被引量:5
  • 4Sonka M,Hlavac V,Boyle R.Image processing,analysis,and machine vision[M].(3rd Edition).Toronto:Thomson,2008:197-207.
  • 5Martelli A.Edge detection using heuristic search methods[J].Computer Graphics and Image Processing,1972,1 (2):169-182.
  • 6Sonka M,Zhang Xiangmin,Siebes M,et al.Segmentation of intravascular ultrasound images:A knowledge-based approach[J].IEEE Transactions on Medical Imaging,1995,14 (4):719-732.
  • 7曲怀敬,孙丰荣,李艳玲,刘泽,宫延新,张梅.基于活动轮廓模型和统计特征的血管内超声图像的边缘提取[J].中国图象图形学报,2005,10(8):999-1004. 被引量:10
  • 8Bovenkamp EGP,Dijkstra J,Bosch JG,et al. Multi-agent segmentation of IVUS images[J].Pattern Recognition,2004,37(4):647-663.
  • 9Bovenkamp EGP,Dijkstra J,Bosch JG.,et al. User-Agent Cooperation in Multiagent IVUS Image Segmentation[J].IEEE Transactions on Medical Imaging,2009,28(1):94-105.
  • 10MacQueen J. Some methods for classification and analysis of multivariate observations [C]// Proceedings of the fifth Berkeley Symposium on Mathematical Statistics and Probability.Berkeley:University of California Press,1967:281-297.

二级参考文献40

  • 1曲怀敬,孙丰荣,李艳玲,刘泽,宫延新,张梅.基于活动轮廓模型和统计特征的血管内超声图像的边缘提取[J].中国图象图形学报,2005,10(8):999-1004. 被引量:10
  • 2Plissiti M E, Fotiadis D I, Michalis L K, et al. An automated method for lumen and media-adventitia border detection in a sequence of IVUS frames [J]. IEEE Transactions on Information Technology in Biomedicine, 2004, 8(2): 131 - 141.
  • 3Sonka M, Zhang X M, Siebes M, et al. Segmentation of intravascular ultrasound images: A knowledge-based approach [J]. IEEE Transactions on Medical Imaging, 1995, 14 (4) : 719 - 732.
  • 4Herrington D M, Johnson T, Santago P, et al. Semi-automated boundary detection for intravascular ultrasound [A]. In : IEEE Proceedings of Computer in Cardiology [C]. Durham, NC, USA,1992:103 - 106.
  • 5Brusseau E, de Korte C L, Mastik F, et al. Fully automatic luminal contour segmentation in intracoronary ultrasound imaging-A statistical approach [J]. IEEE Transactions on Medical Imaging,2004,23 (5) :554 - 566.
  • 6Kass M , Witkin A , Terzopoulos D. Snakes: Active contour models[J]. International Journal of Computer Vision, 1988, 1 ( 4 ) :321 -331.
  • 7Amini A A, Weymouth T E, Jain R C. Using dynamic programming for solving variational problems in vision [J]. IEEE Transactions on Pattern Analysis and Machine Intelligence, 1990,12 (9) : 855 - 867.
  • 8Zhu H, Liang Y, Friedman M H. IVUS image segmentation based on contrast [A]. In: Processing of SPIE[C]. Durham, NC, USA,2002, 4684:1727 - 1733.
  • 9Loupas T, Mcdicken W N, Allan P L. An adaptive weighted median filter for speckle suppression in medical ultrasonic images [J]. IEEE Transactions on Circuits and Systems, 1989, 36( 1 ) : 129 - 135.
  • 10Herrington DM. Semi-automated boundary detection for intravascular ultrasound[A]. In: Herrington DM, Johnson T, Santago P, et al.Computer in Cardiology[C]. Los Alamitos, 1992,103 - 106.

共引文献15

同被引文献71

  • 1唐朝枢,齐永芬.关注血管钙化发病新机制的研究[J].中国医学前沿杂志(电子版),2010,2(3):5-8. 被引量:10
  • 2齐永芬.关注血管钙化的基础和临床研究[J].中国动脉硬化杂志,2015,23(5):433-436. 被引量:14
  • 3Marianne JS,Aaron F.Measuring flow-mediated dilation through transverse and longitudinal imaging:comparison and validation of methods[J].Physics in Medicine and Biology,2010,55(21):6501.
  • 4Vogel RA.Coronary risk factors,endothelial function,and atherosclerosis:a review[J].Clinical Cardiology,1997,20(5):426-432.
  • 5Korkmaz H,Onalan O.Evaluation of endothelial dysfunction:flow-mediated dilation[J].Endothelium,2008,15(4):157-163.
  • 6Thijssen DH,Black MA,Pyke KE,et al.Assessment of flowmediated dilation in humans:a methodological and physiological guideline[J].American Journal of Physiology-Heart and Circulatory Physiology,2011,300(1):H2-H 12.
  • 7Nakamura T,Obata J-e,Hirano M,et al.Endothelial vasomotor dysfunction in the brachial artery predicts the short-term development of early stage renal dysfunction in patients with coronary artery disease[J].International Journal of Cardiology,2011,148(2):183-188.
  • 8Chaudhry FA,Bangalore S,Upadya S,et al.Cross-sectional imaging identifies flow-mediated vasodilatation more accurately compared with longitudinal imaging[J].Journal of the American Society of Echocardiography,2007,20(12):1380-1385.
  • 9Potter K,Reed CJ,Green DJ,et al.Ultrasound settings significantly alter arterial lumen and wall thickness measurements[J].Cardiovascular Ultrasound,2008,6(1):6.
  • 10Zahalka A,Fenster A.An automated segmentation method for three-dimensional carotid ultrasound images[J].Physics in Medicine and Biology,2001,46(4):1321-1342.