摘要
我院供应室针对使用过的针头存在HBsAg和热原污染 ,对针头清洗程序不断改进。采用方法 :(1 )0 1 %新洁尔灭预消毒后一般清洗加清水超声清洗 ;(2 ) 0 5 %消毒灵预消毒后常规操作加清水超声清洗 ;(3) 2 %戊二醛预消毒后常规操作 (煮沸用 1 %碳酸氢钠水 )加清水超声清洗 ;(4) 2 %戊二醛预消毒后常规操作 (煮沸用 1 %碳酸氢钠水 )加清水超声清洗再干烤四种方法进行洗涤质量对比。结果显示 :c组、d组洗涤质量最好。高压灭菌后 ,热原阴性率均达到 1 0 0 %
Due to used needle contaminated by HBsAg and pyrogen we continuously improve our needle washing procedure in supple room of our hospital. Fou kinde of methods were adopted:(1) general washing added ultrasonic washing with clean water after predisinfection by use of 0.1% bromogeramine;(2)routine operation added ultrasonic washing with clean water after predisinfection of 0.5% xiao du ling;(3)routine operation (boiling 1% sodium bicarbonate solution) added ultrasonic washing with clean water after predisinfection of 2% glutaraldehyde. (4)routine operation (boiling 1% sodium bicarbonate solution) added ultrasonic washing with clean water and baking procedure. Washing quality of four kinds of methods mentioned above had been compared. Results showed that:the washing quality of group C and group D were better than that of other methods. After high pressure sterilization, the negative rate of pyrogen were all 100%.