摘要
The inertias of the fluid and the inner cylinder in coaxial cylinder rheometer (CACR) have great influence on the unsteady flow of non-Newtonian fluid. Even for the Newtonian fluid there exist the so called stress overshoot phenomenon. This phenomenon was studied and a method correcting the measured results for an unsteady flow in the rheometer was proposed. It is found that the inertia effect of the fluid can be ignored when the gap between cylinders is small. (Edited author abstract) 11 Refs.
The inertias of the fluid and the inner cylinder in coaxial cylinder rheometer (CACR) have great influence on the unsteady flow of non-Newtonian fluid. Even for the Newtonian fluid there exist the so called stress overshoot phenomenon. This phenomenon was studied and a method correcting the measured results for an unsteady flow in the rheometer was proposed. It is found that the inertia effect of the fluid can be ignored when the gap between cylinders is small. (Edited author abstract) 11 Refs.
基金
The project was supported by the Youth Foundation of the Fourth Military Medical University