期刊文献+

地震传感器的量值溯源方法 被引量:1

Method of value traceability for seismic transducers
在线阅读 下载PDF
导出
摘要 地震传感器在地震学研究与防震减灾领域应用广泛,数量众多,将其测量量值溯源到国际单位,能够保证测量结果准确和统一。文中对地震传感器测量量值进行量值溯源的必要性进行了叙述,说明了地震传感器的量值单位与国际标准单位的关系。对地震传感器量值溯源的途径以及与校准之间的联系进行了阐述。以力平衡加速度计为例对地震传感器的工作原理进行了说明。参照有关的国际标准,研究了基于正弦逼近法的地震传感器绝对法校准的技术要求,并给出了量值溯源框图,为建立地震传感器的量值溯源体系提供了技术依据。最后对低频振动校准的技术发展方向也作了阐述。 Large number of seismic sensors are applied and used widely in seismology research and earthquake disaster prevention. In order to ensure the precieness and unity of measurement results, the measuring value shall be traced to international system of units(SI). The necessity of traceability for measurement result of seismic meters is described in this paper. The relationships between value units of seismic meters and international system of units are explained. The connections of traceability approach for seismic meters and calibrations are elaborated. Working principle of seismic meters was described in example of force balance accelerometer. Refering to the concerning in- ernation standards, technical requirements for absolute calibration of seismic meters based on sine-approximation method are researched. A traceability block diagram is given in this paper. All the research provides a technology basis for seismic sensors value traceability system. In the final, the direction for technical development of low frequency vibration calibration is elaborated also.
出处 《世界地震工程》 CSCD 北大核心 2011年第3期157-162,共6页 World Earthquake Engineering
基金 地震行业科研专项项目(200708040)
关键词 地震传感器 国际标准单位 量值溯源 绝对校准 seismic meters international system of units (SI) value traceability absolute calibration
  • 相关文献

二级参考文献12

  • 1孙桥,于梅.振动测量中正弦逼近法的软件实现及研究[J].中国计量,2004(10):73-74. 被引量:7
  • 2于梅,孙桥,冯源,马明德.正弦逼近法振动传感器幅相特性测量技术的研究[J].计量学报,2004,25(4):344-348. 被引量:26
  • 3于梅,孙桥.外差式激光干涉仪应用于正弦直线和旋转振动测量技术的研究[J].计量学报,2005,26(3):237-241. 被引量:17
  • 4Sill R D. A quadrature laser interferometer for accelerometer calibraion[D]. Endevco, TP 316, 2000.
  • 5ISO 16063 - 11. Methods for the calibration of vibration and shock transducers - Part 11: Primary vibration calibration by laser interferometry (third edition) [ S].
  • 6Tribolet J M. A new phase unwrapping algorithm[J]. IEEE Transactions on Acoustics, Speech, and Signal Processing,1977, 25(2): 170 - 177.
  • 7Heydeman P L M. Determination and correction of quadrature fringe measurement errors in interferometers [ J ]. Applied Optics, 1981, 20(19): 3382 - 3384.
  • 8ISO 1606311,Methods for the calibration of vibration and shock transducers-Part 11:Primary vibration calibration by laser interferometry (third edition)[S].
  • 9SUN Qiao.A Novel Implementation of Homodyne Time Interval Analysis Method with Ordinary Michelson Interferometer for Primary Vibration Calibration Proc.Of the 9th WESPAC Conference,2006.
  • 10Von Martens H-J.Traceability of vibration and shock measurements by laser interferometry[J].Measurement,2000,28:3-20.

共引文献108