期刊文献+

两种不同方法监测中心静脉压的研究 被引量:3

The study of monitoring central venous pressure with two different methods
原文传递
导出
摘要 目的 探讨传统的简易中心静脉压测定法和通过压力传感器监测中心静脉压两种方法的差异性,为临床推广应用提供依据.方法 对我院2008年6月至2009年9月150例实施体外循环心脏直视术患者术前、术后同时采用两种方法测定的中心静脉压值进行比较.结果 两种方法测量所得的中心静脉压值差异无统计学意义(P>0.05).结论 简易中心静脉测压方法既方便又直观,能提高抢救成功率,在基层医院值得推广.压力传感器可连续、动态监测中心静脉压,节省时间,减少护理工作量,便于在大中型医院推广使用. Objective To explore the difference of the traditional simple measurement of central venous pressure and pressure sensor measurement of central venous pressure.Methods The values were compared which were measured by two methods before and after open heart surgery in 150 cases from June 2008 to September 2009.Results The central venous pressure value measured by two methods had no statistically difference(P>0.05).Conclusions Simple measurement method of central venous pressure is convenient and intuitive,which can improve the success rate and in primary hospitals should be promoted.Pressure sensors can continuously,dynamically monitor central venous pressure,save time,reduce nursing workload. It should be promoted in the medium and large hospitals.
出处 《国际护理学杂志》 2011年第7期974-976,共3页 international journal of nursing
关键词 中心静脉压 监测 简易测定法 压力传感器测定法 Central venous pressure Monitor Simple measurement Pressure sensor measurement
  • 相关文献

参考文献6

二级参考文献5

共引文献86

同被引文献23

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部