定做式人工髋关节的CAM实践分析 被引量:1

The Analysis of CAM Implementation of Custom Artificial Hip Joint
在线阅读 下载PDF
导出
摘要 本文介绍定做式人工髋关节成功实施过程。首先对人工髋关节结构进行介绍,然后论述在CADSS环境下人工髋关节这样一个三维曲面CAM的具体实施过程,对刀具的选取、加工余量刀具轨迹等工艺参数进行详细分析。对于加工中碰到的过切问题提出了“保护罩”的解决方案。最后结合X光片进行了术后分析。 in This paper presents how successfully the custom artificial hip joint is implemented. First of all, the structure of custom artificial hip joint is introduced. Then,in the environment of CADDS5, the paper discusses the concrete steps of the implementation of artificial hip joint,which is a 3-D surface. And it is analyzed how to select cutter as well as process parameters like stock. For the problem of overcut in operation, the solution, 'protecting cover', is provided. Finally, the post surgery analysis is performed with the help of X radiograph.
作者 王涛 蔡建国
机构地区 上海交通大学
出处 《机械设计与研究》 CSCD 北大核心 1999年第4期26-28,共3页 Machine Design And Research
关键词 定做式 人工髓关节 三维曲面 CAM Custom artificial hip joint CADDS5 3-D surface 'Protecting cover
  • 相关文献

参考文献2

  • 1凌玲,博士学位论文,1997年
  • 2上海交通大学中德CIMS培训中心,MAHO600 CMAHO CNC532手册

同被引文献14

  • 1连平,夏莉娜,孙荣华,杨维权,尹志强,胡庆沈,贾连顺.腰骶椎管及其内容物三维图像重建的实验研究[J].中华外科杂志,1996,34(6):330-332. 被引量:7
  • 2Bowman SM,Guo XE,Cheng DW, et al. Creep contributes to the fague behavior of bovine trabecular bone. J Biomech, 1998,120(5):647.
  • 3Martini F, Sell S, Kremling E, et al.Determination of periprosthetic bone density with the DEXA method after implantation of custom-made uncemented femoral stems. Int Orthop, 1996,20(4) : 218.
  • 4Dujardin FH, Mollard R, Toupin JM,et al. Micromotion, fit,and fill of custom made femoral stems designed with an automated process. Clin Orthop, 1996,325 : 276.
  • 5Aldinger G, Weipert A. 3D-based production of hip joints : the Aldinger system. Radioloze, 1991 ; 31 (10) :474.
  • 6Reuben JD, Chang CH, Akin JE,et al. A knowledge-based computer-aided design and manufacturing system for total hip replacement. Clin Orthop, 1992,285:48-56.
  • 7Clarke H J, Jinnah RH, Cox QG, et al. Computerized templating in uncemented total hip arthroplasty to assess component fit and fill. J Arthroplasty, 1992; 7:235-239.
  • 8Ioannis DG, Theodore AX, Michael H. et al. Threediemensional computerized selection of hip prostheses in patients with congenital dislocated hips, Orthopedics, 2001,11:269.
  • 9Viceconti M,Testi D. A new hybrid environment for the design of custom-made hip prothesis. Comput Methods Programs Biomed, 2001,64(2):137.
  • 10Reuben JD,Chang CH,Akin JE. A knowledge based computeraided design and manufacturing system for total hip replacement. Clin Orthop, 1992; 258:48.

引证文献1