摘要
目的探讨BC5200血液细胞分析仪全血和稀释模式结果一致性。方法比对两种模式不同分布范围(高、中、低值)血液细胞分析参数。结果稀释模式低值WBC、RBC和高值HGB、MCV结果与全血模式差异有统计学意义(P<0.05),任何范围的PLT测量结果比较差异有统计学意义(P<0.01)。结论 BC5200血液细胞分析仪稀释和全血模式分析结果存在一定的差异,条件许可时应避免使用稀释模式。
Objective To study the consistency of test results using BC5200 hematology analyzer between whole blood and diluent models.Methods We compared the blood cell parameters in different ranges(high,medium,low)between whole blood system and diluent system.Results There was a significant difference(P0.05)of the low content of WBC,RBC and high content of HGB,MCV in the diluent and in whole blood models,and there were significant differences in test results with any range of PLT(P0.01).Conclusion Between diluent and whole blood modes there are significant differences in the results of BC5200 hematology analyzer.Thus,we do not advocate to use the dilution model even if the conditions permit.
出处
《检验医学与临床》
CAS
2011年第8期916-917,共2页
Laboratory Medicine and Clinic
关键词
血液细胞分析仪
稀释
临床应用
hematology analyzer
diluent
clinical application