期刊文献+

Characteristics of the magnetic analysis system for a compact MPR-type spectrometer

Characteristics of the magnetic analysis system for a compact MPR-type spectrometer
原文传递
导出
摘要 The magnetic proton recoil(MPR)spectrometer is a novel diagnostic instrument with high perfor-mance for measurements of neutron spectra in inertial confinement fusion(ICF)experiments and high power fusion devices.A compact MPR-type spectrometer dedicated to the research of pulsed deuterium-tritium(DT)neutron spectroscopy of special experimental conditions is currently under design.Analyses of the main parameters and performance of the magnetic analysis system through 3-D particle transport calculations and MonteCarlo simulations and calibration of the system performance as a test using CR-39 solid track detector and α particle from 239pu and 226Ra radioactive sources are presented in this paper.The results indicate that the magnetic analysis system will achieve a detection efficiency level of 10-5-10-4 at an energy resolution of 1.5%-2.1%,and fulfills the design goals of the spectrometer. The magnetic proton recoil(MPR)spectrometer is a novel diagnostic instrument with high perfor-mance for measurements of neutron spectra in inertial confinement fusion(ICF)experiments and high power fusion devices.A compact MPR-type spectrometer dedicated to the research of pulsed deuterium-tritium(DT)neutron spectroscopy of special experimental conditions is currently under design.Analyses of the main parameters and performance of the magnetic analysis system through 3-D particle transport calculations and MonteCarlo simulations and calibration of the system performance as a test using CR-39 solid track detector and α particle from 239pu and 226Ra radioactive sources are presented in this paper.The results indicate that the magnetic analysis system will achieve a detection efficiency level of 10-5-10-4 at an energy resolution of 1.5%-2.1%,and fulfills the design goals of the spectrometer.
出处 《Chinese Physics C》 SCIE CAS CSCD 2010年第12期1860-1865,共6页 中国物理C(英文版)
基金 Supported by Science and Technology Development Foundation of China Academy of Engineering Physics (2008B0103003)
关键词 fusion neutron magnetic proton recoil (MPR) magnetic analysis SPECTROMETER fusion neutron, magnetic proton recoil (MPR), magnetic analysis, spectrometer
  • 相关文献

参考文献9

  • 1Jarvis O N.Nucl.Instrum.Methods in Phys.Res.A,2002,476:474-484.
  • 2K(a)llne Jet al.Rev.Sci.Instrum.,1999,70(1):1181-1184.
  • 3CHEN M et al.Nuclear Electronics & Detection Technology,2002,22(5):462-464.
  • 4Aymar R.Fusion Engineering and Design,2002,61-62:5-12.
  • 5Sj(o)strand H et al.Rev.Sci.Inetrum.,2006,77:10E717.
  • 6Maas A C et al.Fusion Engineering and Design,1999,47:247-255.
  • 7Frenje J et al.Rev.Sci.Instrum.,1999,70(1):1176-1180.
  • 8Frenje J et al.Rev.Sci.Instrum.,2001,72(1):854-858.
  • 9Nishitani T et al.Rev.Sci.Instrum.,1997,68(1):565-568.