期刊文献+

SysmexXT-1800i全自动血细胞分析仪性能评价 被引量:8

An Evaluation on Performance of SysmexXT-1800i Automated Hematology Analyzer
暂未订购
导出
摘要 目的:评价SysmexXT-1800i全自动血细胞分析仪的性能。方法:对SysmexXT-1800i的精密度、携带污染率、线性测定、总重复性进行测定,并将其白细胞分类计数(DC)结果与手工分类结果进行比较。结果:Sys-mexXT-1800i的精密度、携带污染率、线性测定及总重复性等指标数值都在允许范围内;与手工法DC结果比较,中性粒细胞、淋巴细胞、单核细胞、嗜酸性粒细胞和嗜碱性粒细胞相关系数(r)分别为0.96、0.954、0.462、0.807和0.618。结论:SysmexXT-1800i全自动血细胞分析仪主要性能指标符合实验要求,是较理想的全自动血细胞分析仪。 Objective:To evaluate the function of SysmexXT-1800i automated hematology analyzer. Methods:To examine the precision,carry over rate,linearity and total repeatability of SysmexXT-1800i,then compare its differential counts (DCs) of white blood cells with that obtained manually. Results:The precision,carryover rate,linearity and total repeatability were in the prescriptive range,and the correlation coefficients (r) between DCs of neutrophils,lymphocytes,monocytes,eosinophils,and basophils obtained by handiwork and the machine were 0.96,0.954,0.462,0.807,and 0.618,respectively. Conclusion:The main performance indexes of SysmexXT-1800i accord with the experimental requirements,which shows the machine is an ideal automated hematology analyzer.
出处 《贵阳医学院学报》 CAS 2010年第3期275-277,282,共4页 Journal of Guiyang Medical College
关键词 白细胞计数 红细胞计数 血小板计数 血细胞分析仪 leukocyte count erythrocyte count platelet count hematology analyzer
  • 相关文献

参考文献3

二级参考文献8

  • 1International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analyzer including those used for differential leukocyte and reticulocyte counting and cell marker application. Prepared by the ICSH expert panel on cytometry [ J ].Clin Lab Haemat, 1994,16:157-174.
  • 2National Committee for Clinical Laboratory Standards. Leukocyte Differential Counting NCCLS Approved Standard H20-A2 [ S ].Villanova, PA, 1992.
  • 3Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure [ J ]. Br J Haematol, 2000, 108:228-235.
  • 4Hallawell R,O Malley C,Hussein S,et al. An evaluation of the Sysmex NE-8000 hematology analyzer[J]. Am J Clin Pathol ,1991,96 :594-601.
  • 5Carter A B, Chavda N, Grant D,et al. Evaluation of the Sysmex SE-9000 automated hematology analyzer[ J]. Sysmex J Int, 1995,5:6-16.
  • 6Vives-Corrons J L,Besson I,Jou J M,et al. Evalution of the Abbott Cell-DYN 3500 hematology analyzer in a university hospital[ J]. Am J Clin Pathol, 1996,105:553-559.
  • 7Thalhammer-Seherrer R, Knobl P, Korninger L, et al. Automated fivepart white blood cell differential counts, efficiency of softwaregenerated white blood cell suspect flags of the hematology analyzers Sysmex SE-9000, Sysmex NE-8000, and Coulter STKS [ J ]. Arch Pathol Lab Med, 1997,121:573-577.
  • 8Bentley S A, Johnson A, Bishop C A. A parallel evaluation of four automated hematology analyzers [ J ]. Am J Clin Pathol, 1993,100:626-632.

共引文献16

同被引文献71

  • 1王文娟,王佩佩,陈保德,王磊,钟步云.LH750血液分析仪临床应用评价[J].中华检验医学杂志,2005,28(3):319-321. 被引量:32
  • 2张莉,吴炯,郭玮,潘柏申.医学检验检测系统应用前的性能评价[J].检验医学,2006,21(6):560-563. 被引量:71
  • 3杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.
  • 4ICSH. Guidelines for the evaluation of blood cell analysers inclu-ding those used for differential leucocyte and reticulocyte countingand cell marker application[J]. Clin Lab Haemat, 1994,16(2):157-174.
  • 5CLSI. EP9-A Method comparison and bias estimation using pa-tient sample; Aproved guidelines[S]. Wayne, PA: CLSI, 1986 . 1-14.
  • 6丛玉隆,岳家新,袁家颖.实用血细胞分析技术与临床[M].北京:人民军医出版社,2011:14.
  • 7蒋秉坤,范钦信.临床生物化学及生物化学检验[M].4版.北京:人民卫生出版社,2001:15-28.
  • 8迈瑞公司.迈瑞全自动血液细胞分析仪5180用户使用说明书[M].深圳:深圳迈瑞电子股份有限公司:25-36.
  • 9冯仁丰.临床实验室建议与管理[M].上海:上海科学技术出版社,1991:127-170.
  • 10International Council for Standardization in Hematology. Guidelines for the evhluation of blood cell analyzer including those used for differential leukocyte and reticulocyte counting and cell marker application[J]. Clin LabHaematol,1994(2):157-174.

引证文献8

二级引证文献40