自动毫安低管电压技术在头颈部血管造影中应用的可行性研究 被引量:18

The application of anto-mA low-kV CT scanning on head and neck CT angiography: a feasibility study
暂未订购
导出
摘要 目的:评估自动毫安低管电压64层螺旋CT扫描对头颈部血管造影图像质量及辐射剂量的影响。方法:收集60例行头颈CTA检查的患者。随机分为4组,Ⅰ组管电压为100kV,噪声指数(NI)为20;Ⅱ组管电压为120kV,NI为20;Ⅲ组管电压为100kV,NI为10;Ⅳ组管电压为120kV,NI为10。使用64层螺旋CT进行扫描。定量和定性分析图像质量,并记录辐射剂量。统计学分析采用单因素方差分析,P<0.05为差异有统计学意义。结果:动脉CT值、图像噪声、背景信号的结果有统计学差异(P<0.05)。在NI相同时,kV值降低,动脉CT值升高,噪声升高,背景也升高;而信噪比(SNR)和对比噪声比(CNR)无统计学差异。NI相同时,管电压为120kV组与100kV组的图像质量评分无差别,管电压相同时,NI=10组图像质量优于NI=20组。结论:自动毫安低剂量CT扫描技术可以在明显降低辐射剂量的同时确保图像质量,管电压为100kV、NI为20可以满足临床诊断需要。 Objective: To evaluate the effect of auto-mA low-kV CT scanning on image quality and radiation dose for the head and neck CT angiography. Methods: This prospective study was performed in 60 patients needed head and neck CTA. Patients were randomly divided into four groups: group I kV=100, NI=20; group II : kV=120, NI=20; group III: kV=100, NI=10; group IV kV=120, NI=10. Images were obtained with 64 mutislice spiral CT scanner(GE LightSpeed VCT XT). The other parameters include: slice thickness 0.625mm, pitch 0.984, 0.4s/rot, FOV 22cm, matrix 512*512, standard algorithm. Images were all transmitted to off-line workstation (GE AW4.4) and volume rendering (VR), muhiplanar reformat (MPR) were reformatted for further scoring. Quantitative and qualitative analysis of image were performed, and radiation dose was recorded. The methods of statistics analysis were ANOVA, and P〈0.05 represented the significant difference. Results: The differences between CT value, noise and background signal were significant. In the same NI, with decreasement of kV, CT value, noise, and background signal were all increased. However, SNR and CNR had no significant difference,. In the same NI, there was no significant difference among the image quality of four groups. In the same kV, the image quality of two NI=10 groups was obvious bettter than the other two NI=20 groups. Conclusion: Auto-mA low-dose CT scanning can reduce radiation dose significantly, meanwhile ensure the image quality. Scan protocol of kV=100 and NI= 20 can guarantee the imaging quality of diagnostic purpose.
出处 《天津医科大学学报》 2010年第2期322-325,共4页 Journal of Tianjin Medical University
关键词 体层摄影术 X线计算机 动脉血管造影 自动毫安技术 低剂量 Tomography, x-ray computed Artery angiography Auto-mA Low-dose
  • 相关文献

参考文献10

  • 1Macari M,Bini EJ,Xue X,et al.Colorectal neoplasms:prospective comparison of thin-section low-dose multi-detector row CT colonography and conventional colonoscopy for detection[J].Radiology,2002,224(2):383.
  • 2Bond JH.Screening guidelines for colorectal cancer[J].AM J Med,1999,106(1A):7S.
  • 3王雪飚,张桂池,郭德安.16层螺旋CT超低剂量仿真结肠镜及其后处理诊断结肠疾病的价值[J].放射学实践,2006,21(3):276-280. 被引量:20
  • 4Campbell J,Kalra MK,Rizzo S,et al.Scanning beyond anatomic limits of the thorax in chest CT:findings,radiation dose,and automatic tube current modulation[J].AJR,2005,185(6):1525.
  • 5Herzog C,Mulvihill DM,Nguyen SA,et al.Pediatric cardiovascular CT angiography:radiation dose reduction using automatic tube current modulation[J].AJR,2008,190(5):1232.
  • 6Husmann L,Valenta I,Gaemperli O,et al.Feasibility of low-dose coronary CT angiography:first experience with prospective ECGgating[J].Eur Heart J,2008,29(2):191.
  • 7任庆云,赵增仁,李敬玉.腹部低剂量CT扫描的临床研究[J].中国辐射卫生,2006,15(4):495-496. 被引量:11
  • 8Szucs-Farkas Z,Kurmann L,Strautz T,et al.Patient exposure and image quality of low-dose pulmonary computed tomography angiography:comparison of 100-and 80-kVp Protocols[J].Invest Radiol,2008,43(12):871.
  • 9Keller MR,Kessler RM,Brooks RA,et al.Optimum energy for performing CT iodinated contrast studies[J].Br J Radiol,1980,53(630):576.
  • 10Huda W,Scalzetti EM,Levin G.Technique factors and image quality as functions of patient weight at abdominal CT[J].Radiology,2000,217(2):430.

二级参考文献20

  • 1Greelee RT,Murray T,Bolden S,et al.Cancer Statistics[J].CA Cancer J Clin,2000,50(1):7-33.
  • 2Hara AK,Johnson CD,Maclarty RL,et al.CT Colonography:Single-versus Multi-detector Row Imaging[J].Radiology,2001,219(2):461-465.
  • 3Dachman AH,Yoshida H.Virtual Colonography:Past,Present and Future[J].Radiol Clin N Am,2003,41(2):377-393.
  • 4Fischbach R,Wessling J.CT Colonography Using Multi-slice Computed Tomography[J].Electromedica,2001,69 (2):116-119.
  • 5Van GRE,Venema HW,Serlie IWO,et al.CT Colonography at Different Radiation Dose Levels:Feasibility of Dose Reduction[J].Radiology,2002,224 (1):25-33.
  • 6Yee J,Akerkar GA,Hung RK,et al.Colorectal Neoplasia:Performance Characteristics of CT Colonography for Detection in 300Patients[J].Radiology,2001,219(3):685-692.
  • 7Iannaccone R,Laghi A,Catalano C,et al.Feasibility of Ultra-lowdose Multislice CT Colonography for the Detection of Colorectal Lesions:Preliminary Experience[J].Eur Radiol,2003,13 (10):1297-1302.
  • 8Beaulieu CF,Jeffrey RB,Karadi C.Display Modes for CT Colonography Part Ⅱ Blinded Comparison of axial CT and Virtual Endoscopic and Panoramic Endoscopic Volume-rendered Studies[J].Radiology,1999,212(1):203-212.
  • 9Kalra MK,Prasad S,Saini S,et al.Clinical comparision of standard-dose and 50% reduced-dose abdominal CT:effect on image quality[J].AJR,2002,179-1101
  • 10Hopper KD,Keeton NC,Kasales CJ,et al.Utility of low mAs 1.5 pitch helical versus conventional high mA abdominal CT[J].Clinical Image,1998,22:54-59

共引文献28

同被引文献177

引证文献18

二级引证文献129

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部