参加肌酸激酶参考方法国际室间比对计划结果分析 被引量:7

Analysis of the results of creatine kinase reference method in ring trial program
原文传递
导出
摘要 目的 对2006-2008年3年来参加IFCC组织的参考实验室RELA的CK的测定结果 进行比较和分析,为我国CK参考方法 的建立和应用提供借鉴.方法 按照IFCC公布的酶学活性测定(37℃)参考方法 的标准操作程序(SOP)测定CK国际样本,不精密度评价按美国CLSI的EP5-A2进行,准确度评价采用国际有证参考物质(ERM).结果 2006年测定结果 ,样本A CK值为(9.896±0.112)μkat/L,样本B为(4.953±0.120)μkat/L;2007年测定结果 ,样本A为(2.684 ±0.054)μkat/L,样本B为(8.798 ±0.101)μkat/L;2008年测定结果 ,样本A为(10.523±0.149)μkat/L,样本B为(10.551±0.141)μkat/L.2006-2008年CK参考方法 的不精密度分别为0.92%、0.86%、0.88%,均<1%,ERM测定均值在"靶值±不确定度[(1.68±0.07)μkat/L]"范围内,初步验证了参考方法 的不精密度和准确度.结论 连续3年的CK国际样本测定结果 均在IFCC公布的置信区间(limit of equivalence)±5%范围内,CK的IFCC参考方法 已经建立并且日趋成熟. ObjectiveTo provide a reference for the establishment and appliance of enzyme reference measurement of CK in China by comparing and analyzing the RELA results of CK in IFCC from 2006-2008. Methods The RELA samples of CK were measured according to the reference procedure for the measurement of catalytic activity concentration of CK (37 ℃) ,which had been published by IFCC. The EP5A2 protocol was used for evaluation of the imprecision and ERM was used for verification of the trueness. Results In RELA 2006, the result of sample A was (9. 896 ±0. 112) μkat/L, and the result of sample B was (4.953 ±0. 120) μkat/L. In RELA 2007, the result of sample A was (2.684 ±0.054)μkat/L, and the result of sample B was (8.798 μ0. 101) μkat/L. In RELA 2008, the result of sample A was (10. 523 ±0. 149) μkat/L,and the result of sample B was (10. 551 ±0. 141) μkat/L. The precision of the CK reference method in the year 2006 to 2008 was 0. 92%, 0. 86% and 0. 88% respectively, each of them is less than 1% and the results of ERMs were consistent with the certified value(1. 68 ± 0. 07)μkat/L,which verify the imprecision and accuracy of the reference method. Conclusions All of the results in the continuous three years were in the range of equivalence limits suggested by IFCC. The CK reference method suggested by IFCC has been established and it is getting better.
出处 《中华检验医学杂志》 CAS CSCD 北大核心 2010年第6期554-558,共5页 Chinese Journal of Laboratory Medicine
基金 首发基金重大联合项目:北京市临床生化参考测量体系在社区的推广应用(2007-390)
关键词 肌酸激酶 参考标准 评价研究 临床酶试验 Creatine kinase Reference standards Evaluation studies Clinical enzyme tests
  • 相关文献

参考文献11

  • 1Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Part 2.Reference procedure for the measurement of catalytic activity concentration of ereatine kinase.Clin Chem Lab Med,2002,40:635-642.
  • 2Schumann G,Bonora R,Cefiotfi F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Part 3.Reference procedure for the measurement of catalytic activity concentration of lactate dehydrogenase.Clin Chem Lab Med,2002,40:642-648.
  • 3Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Part 4.Reference procedure for the measurement of catalytic activity eoneentration of alanine aminotransferase.Clin Chem Lab Med,2002,40:718-724.
  • 4Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Part 5.Reference procedure for the measurement of catalytic activity concentration of aspartate aminotransferase.Clin Chem Lab Med,2002,40:725-733.
  • 5Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Part 6.Reference procedure for the measurement of catalytic activity concentration of γ-glutamyltransferase.Clin Chem Lab Med,2002,40:734-738.
  • 6Schumann G,Aoki R,Ferrero CA,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.Clin Chem Lab Med,2006,44:1146-1155.
  • 7International Organization for Standardization.ISO 18153 In vitro diagnostic medical device-Measurement of quantities in samples of biological origin-Metrologieal traceability of values for catalytic concentration of enzyme assigned to calibrators and control materials.Geneva:ISO,2003.
  • 8金丕焕.医学统计学.上海:上海医科大学出版社,1993.
  • 9Clinical and Laboratory Standards Institute.EPS-A2 Evaluation of precision performance of quantative measurement methods;approved guideline-second edition.Wayne,PA:CLSI,2004.
  • 10杨振华.中华医学会检验学会酶活性浓度测定的推荐方法,测定人血清(血浆)中酶活性浓度方法总则(草案).中华医学检验杂志,1994,17(2):121-123.

二级参考文献9

  • 1International Organization for Standardization ISO18153. In vitro diagnostic medical device-Measurement of quantities in samples of biological origin-Metrological traceability of values for catalytic concentration of enzyme assigned to calibtators and control materials. 2003.
  • 2Schumann G, Bonora R, Cefiotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 2. Reference procedure for the measurement of catalytic activity concentration of creatine kinase. Clin Chem Lab Med, 2002,40: 635-642.
  • 3Schumann G, Bonora R, Cefiotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 3. Reference procedure for the measurement of catalytic activity concentration of lactate dehydrogenase. Clin Chem Lab Med,2002,40: 642-648.
  • 4Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 4. Reference procedure for the measurement of catalytic activity concentration of alanine aminotransferase. Clin Chem Lab Med, 2002,40. 718-724.
  • 5Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 5. Reference procedure for the measurement of catalytic activity concentration of aspartateaminotransferase. Clin Chem Lab Med,2002,40: 725-733.
  • 6Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 6. Reference procedure for the measurement of catalytic activity concentration of γ- glutamyltransferase. Clin Chem Lab Med,2002,40: 734-738.
  • 7Schumann G, Aoki R, Ferrero CA, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part 8. Reference procedure for the measurement of catalytic concentration of alpha-amylase. Clin Chem Lab Med,2006, 44:1146-1155.
  • 8Siekmann L, Doumas BT, Thienpont L, et al. Reference materials and reference measurement systems in laboratory medicine. Network of reference laboratories. Eur J Clin Chem Clin Biochem, 1995, 33:1013-1017.
  • 9Panteghini M, Ceriotti F, Schumann G, et al. Establishing a reference system in clinical enzymology. Clin Chem Lab Med, 2001, 39:795-800.

共引文献27

同被引文献53

  • 1徐国宾,吴南,王清涛.要重视血清酶学测定的标准化工作[J].临床检验杂志,2007,25(3):161-164. 被引量:25
  • 2杨有业,张秀明.临床检验方法学评价.北京:人民卫生出版社,2008.
  • 3Appert-Flory A,Fischer F,Jambou D,et al.Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP.Thromb Res,2007,120:733-743.
  • 4Fischer F,Appert-Flory A,Jambou D,et al.Evaluation of the automated coagulation analyzer Sysmex CA-7000.Thromb Res,2006,117:721-729.
  • 5Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. International Federation of Clinical Chemistry and Laboratory Medicine. Part 6. Reference procedure for the measurement of catalytic concentration of gamma- glutamyltransferase. [ J]. Clin Chem Lab Med, 2002, 40(7) :734-738.
  • 6National Committee for Clinical Laboratory Standards. Evaluation of precision performance of quantative measurement methods; approved guideline [ S ]. 2nd ed. EP5-A2, NCCLS,2004.
  • 7杨振华.中华医学会检验学会酶活性浓度测定的推荐方法,测定人血清(血浆)中酶活性浓度方法总则(草案).中华医学检验杂志,1994,17(2):121-123.
  • 8陈文祥,申子瑜,杨振华.临床检验中的测量不确定度[J].中华检验医学杂志,2007,30(9):967-971. 被引量:31
  • 9Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentra- tions of enzymes at 37 degrees C. Part 5. Reference procedure for the measurement of catalytic concentration of aspartate amin- otransferase [J]. Clin Chem Lab Med, 2002,40(7) : 725 -733.
  • 10National Committee for Clinical Laboratory Standards. EP15 A User demonstration of performance for precision and accuracy approved guideline[S]. Wayne,PA,NCCLS,2001.

引证文献7

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部