期刊文献+

血球分析仪的校准与比对 被引量:6

Correction and contrast of the examined results among different hemocytoanalyzers
暂未订购
导出
摘要 目的探讨不同血球分析仪检测结果的可比性和一致性的方法。方法根据美国临床化学标准化委员会(NCCLS)标准化文件EP9-A的要求,对二台美国雅培CELL-DYN-3700(仪器1)、CELL-DYN-1700(仪器2)与日本希斯美康SYSMEX-XE-2100(仪器3)血球分析仪的检测结果在校准前和校准后分别进行比对分析,以评估两检测系统检测结果的偏倚是否在可接受范围内。结果检测结果表明,校准前美国雅培CELL-DYN与日本希斯美康SYSMEX系统在个别项目上有较大偏倚;而校准后两系的偏倚完全符合要求,且有较好的一致性。结论校准是保证仪器结果稳定的有效方法 ,通过定期校准仪器,能使不同品牌血球分析仪之间达到较好的稳定性及各仪器结果之间的一致性。 Objective An evaluation of the deviation of the examined results and contrast of the methods used by different hemocytoanalyzers was made in order to seek a method to make the results of different instruments more comparable and consistent.Methods According to the requirement of NCCLS,two U.S made hemocytoanalyzers CELL-CYN-3700,CELL-DYN-1700 and one Japan made SYSMES-XE-2100 were compared in their examined results before and after correction to see whether the deviation of the results be acceptable.Results There were great deviations between CELLS-DYN and SYSM-EX in several items before correction.But after correction,the deviation of the examined results by both was acceptable and had a better consistence.Conclusion Correction is a good means to assure a stable result.A better stability and consistence of the results can be achieved by regular correction.
出处 《西部医学》 2010年第6期1110-1112,共3页 Medical Journal of West China
关键词 血球分析仪 校准 比对 Hemocytoanalyzer Correction Contrast
  • 相关文献

参考文献4

二级参考文献2

  • 1National Committee for Clinical Laboratory Standards.Evaluation of matrix effects; Approved guideline[S].NCCLS document EP14-A,2001.
  • 2Mosca A,Paleari R,ScimeDegani V,et al.Intermethod differences and commutability of control material for HbA2 measurement[J].Clin Chem Lab Med,2000,10(38):997-1002.

共引文献59

同被引文献40

  • 1王治国,王薇,李小鹏.测量不确定度及其在临床检验中应用[J].中国卫生统计,2005,22(2):85-86. 被引量:57
  • 2张雯艳,孙庆霞,丁家华.测量不确定度及其在临床检验中的应用[J].中华检验医学杂志,2006,29(7):590-592. 被引量:22
  • 3Jerelyn W,Patricia G.Performance evaluation of the Sysmex XE-2100 hematology analyzer[J].Lab Hemato1,2000,6(2):85-86.
  • 4Kazuto T,Tomohiro T,Tesesuyl U,et al.Evaluation and clinical usefulness of the automatic hematology analyzer.Sysmex XE-2100[J].Sysmex J Int,1999,9(2):129-135.
  • 5Hideaki M.The technology of reagents in the automated hematology analyzer sysmex XE-2100-red fluorescence reaction[J].Sysmex J Int,1999,9(2):179-185.
  • 6Herklotz R,Huber AR.Precision and accuracy of the leukocyte differential in the Sysmex XE-2100[J].Sysmex J Int,2001,11(11):8-21.
  • 7杨有业,张秀明.临床检验方法学评价[M] .北京:人民卫生出版社,2008:64.
  • 8Pilotto A,Franceschi M,Maggi S,et al. Optimal management of pepticulcer disease in the elderly[J]. Drugs and aging,2010;27(7):545-546.
  • 9郭爱芝,闫峰.血细胞分析方法比对探讨[J].实用医技杂志,2008,15(18):2369-2370. 被引量:8
  • 10刘运双,徐鹏,薛冰蓉,曾平,彭国瑞.对比试验在血细胞分析室内质量控制中的应用[J].重庆医学,2009,38(2):180-181. 被引量:9

引证文献6

二级引证文献15