期刊文献+

超声引导经锁骨下静脉置入植入式输液港的效果 被引量:6

Correct placement of the tip of central venous catheter In the vena cava and right atrium through subclavian vein under ultrasound guidance
原文传递
导出
摘要 目的 探讨应用超声引导定位在经锁骨下静脉人路置入植入式输液港的效果.方法 选择乳腺癌患者60例,年龄28~63岁,体重41~70 kg,身高150~168 cm,ASA Ⅰ或Ⅱ级,拟经锁骨下静脉穿刺置入植入式输液港行长期输液及化疗,输液港导管尖端位置应在上腔静脉和右心房交界处.随机分为2组(n=30):对照组(C组)经锁骨下静脉穿刺,采用脉冲注射冰盐水实验的方法引导定位植入式输液港导管尖端位置.超声组(U组)经锁骨下静脉穿刺,采用超声引导定位植入式输液港导管尖端位置.所有患者术后均经过X线摄片进行导管定位.计算两组患者首次置入成功率.结果 U组患者术后X线摄片显示输液港导管尖端均在上腔静脉和右心房交界处,未发生导管偏离进入颈内静脉的现象,首次置入成功率100%.C组术后X线摄片显示有6例患者的输液港导管发生了偏离,进入颈内静脉,需回手术室在超声引导定位下重新穿刺置管,首次置入成功率80%.U组首次置入成功率高于C组(P〈0.05).结论 超声引导定位是经锁骨下静脉正确放置植入式输液港的有效手段. Objeetlve To avoid inserting the tip of the central venous catheter into internal jugular vein (IJV) through subclavian vein under ultrasound guidance.Methods sixty breast cancel patients aged 28-63 yr weighing 41-70 kg who needed long-term intravenous infusion and chemotherapy through central venous catheter were randomly divided into 2 groups (n=30 each):control group (group C) and ultrasound group (group U).In group C the insertion of central venous catheter through subclavian vein was guided by pulsatile injection of ice-cold saline.In group U ultrasound detector (type HFL 38/13-6 MH,Sonosite Co,USA) was used to guide the insertion of tbe central vesons catheter.The position of the catheter tip was verified by X-ray radiography.The rate of successful placement at 1st attempt was calculated.Results The tip of the central venous catheter was correctly placed in the vena cava and right atrium in all patients in group U (success rate 100%),while in group C the tip was misplaced in IJV in 6 patients (success rate 80%) and had to be replaced.Conclusion Ultrasound guidance is effective for correct placement of the tip of central venous catheter in the vema cava and right atrium through subclavian vein.
出处 《中华麻醉学杂志》 CAS CSCD 北大核心 2010年第3期285-287,共3页 Chinese Journal of Anesthesiology
关键词 超声检查 多普勒 锁骨下静脉 输注 静脉内 Ultrasonography,doppler Subclavian vein Infusions,intravenous
  • 相关文献

参考文献6

  • 1郑颖,周子戎,汪春英,张爱萍,闻大翔,黄贞玲,杭燕南.超声引导中心静脉穿刺置管的临床评价[J].上海医学,2007,30(7):504-506. 被引量:22
  • 2Feller-Kopman D.Ultrasound-guided internal jugular access:a proposed standardized approach and implications for training and practice.Chest,2007,132(1):302-309.
  • 3Hind D,Calvert N,McWilliams R,et al.Ultrasonic locating devices for central venous cannulation:meta-analysis.BMJ,2003,327(7411):361.
  • 4Randolph AG,Cook DJ,Gonzales CA,et al.Ultrasound guidance for placement of central venous catheters:a meta-analysis of the literature.Crit Care Med,1996,24(12):2053-2058.
  • 5Atkinson P,Boyle A,Robinson S,et al.Should ultrasound guidance be used for central verous catheterisation in the emergency department?Emerg Med J,2005,22(3):158-164.
  • 6郭瑞军,王明花,于亚东,张经歧,田德虎,张宝珍,张文云,李莉,高石军.高频超声在外周神经、肌腱、肌肉的应用[J].中国超声医学杂志,1998,14(12):32-35. 被引量:28

二级参考文献6

  • 1Sauer W,Luft D,Risler T,et al.Significance of ultrasonics in the placement of a central venous catheter.Dtsch Med Wochenschr,1988,113:1423-1427.
  • 2Sofocleous CT,Schur Ⅰ,Cooper SG,et al.Sonographically guided placement of peripherally inserted central venous catheters:review of 355 procedures.Am J Roentgenol,1998,170:1613-1616.
  • 3Oguzkurt L,Tercan F,Kara G,et al.US-guided placement of temporary internal jugular vein catheters:immediate technical success and complications in normal and high-risk patients.Eur J Radiol,2005,55:125-129.
  • 4Miller AH,Roth BA,Mills TJ,et al.Ultrasound guidance versus the landmark technique for the placement of central venous catheters in the emergency department.Acad Emerg Med,2002,9:800-805.
  • 5Costantino TG,Parikh AK,Satz WA,et al.Ultrasonographyguided peripheral intravenous access versus traditional approaches in patients with difficult intravenous access.Ann Emerg Med,2005,46:456-461.
  • 6Mey U,Glasmacher A,Hahn C,et al.Evaluation of an ultrasound-guided technique for central venous access via the internal jugular vein in 493 patients.Support Care Cancer,2003,11:148-155.

共引文献48

同被引文献23

  • 1Niederhuber JE,Ensminger W,Gyves JW,et al. Totally implan-ted venous and arterial access system to replace external cathetersin cancer treatment[ J]. Surgery,1982,92(4):706-712.
  • 2Ballarini C,Ceretti AP,Cordovana A,et al. Complications of sub-cutaneous infusion post in the general ontology population [ J].Oneology,1999,56(l):97-102.
  • 3Noriaki Sakamoto, Yasuaki Arai, Yoshito Takeuchi, et al. Ultrasound - guided radiological placement of central venous port via the subclavian vein[J ]. Cardiovasc In- tervent Radiol, 2010, 33(5): 989.
  • 4Aldrighetti L, Paganelli M, Arm M, et al. Complications of blind placement technique in 980 subcutaneous infusion ports [J]. J Vase Access, 2000, 1(1): 28.
  • 5陈明远,夏良平,陈直华,吴剑辉,刘巍巍,郭桂芳,谢汝华.植入式静脉输液港不同植入术式在恶性肿瘤患者中的应用[J].中山大学学报(医学科学版),2007,28(B06):145-147. 被引量:38
  • 6Gleson NC,Fiorica JV,Mark JE. Externalized Groshong catheters and Hichman ports for central venous access in gynecologic oncology patients[J].Gynecologic Oncology,1993,(03):372-376.
  • 7Kowalski CM,Kaufman JA,Rivitz SM. Migration of central venous catheters:implications for initial catheter tip positioning[J].J Vase lnterv Radiol,1997,(03):443-447.
  • 8Funaki B. Central venous access:a primer for the diagnostic radiologist[J].American Journal of Roentgenology,2002,(02):309-318.
  • 9戴宏琴,薛嵋.植入式静脉输液港在肿瘤患者中的应用与护理[J].上海护理,2008,8(1):57-58. 被引量:28
  • 10孙凤芝,李永杰,李东,赵莹莹.高频超声引导经锁骨下静脉中心静脉置管术的临床应用价值[J].中国超声医学杂志,2009,25(9):881-884. 被引量:26

引证文献6