摘要
依据OmegaEngineeringHandbook中的钨铼5-钨铼26热电偶分度表和Sand-strom&Withrom提供的钨铼5-钨铼26热电偶在-240~0℃校正数据,对钨铼5-钨铼26热电偶在-240~2300℃进行了多项式拟合,偏差为±1℃。提供了热电势步长为0.1mVW5%RevsW26%Re热电偶的分度表。
The thermocouples of the Tungsten 5% Rhenium Tungsten 26% Rhenium were calibrated from -240 to 2 300 ℃ by polynomial fitting methods,according to the table on the Omega Engineering Handbook and the data for -240 ̄0℃ by Sandstrom & Withrom.The accuracy is ±1 ℃,and a table of W5%Re vs W26%Re by 0.1 mV step is also provided.
出处
《真空电子技术》
1999年第1期11-14,共4页
Vacuum Electronics
关键词
热电偶
多项式拟合法
分度表
Thermocouple,Polynomial fitting methods, Table