摘要
详细介绍了美国Thermo仪器公司生产的Thermo 6300型电感耦合等离子体发射光谱仪在使用过程中不能"引燃"、容易"熄火"、数据稳定性差等常见故障及排除方法,以及其辅助设备的维护维修实例。
This paper describes common faults occured in Thermo 6300 inductively coupled plasma-atomic emission spectrometer running, such as difficult to "ignite", easy to "turn off" and poor data stability. Some troubleshooting examples for instrument and auxiliary equipments were also demonstrated.
出处
《岩矿测试》
CAS
CSCD
北大核心
2010年第2期196-198,共3页
Rock and Mineral Analysis
关键词
电感耦合等离子体发射光谱仪
故障
排除方法
inductively coupled plasma-atomic emission spectrometer (ICP-AES)
fault
troubleshooting