摘要 目的探讨血液分析仪对低值血小板的计数误差。方法将Beckman—CoulterLH-750和SysmexXE-2100全自动血液分析仪检测的120例血小板数≤50×10^9/L的新鲜血液常规标本,同时用血小板计数参考流式细胞术法测定,用Kappa和相关回归等统计学方法分析血液分析仪对低值血小板检测的准确性。结果全自动血液分析仪在血小板计数(0—50)×10^9/L时,与参考方法结果的总符合率均〉90%,Kappa值均〉0.80。但在血小板计数≤15×10^9/L时,与参考方法符合率仅75.0%-82.3%,Kappa值95%的可信限下限均〈0.80。结论全自动血液分析仪对低值血小板计数的准确性较好,但随着血小板数值的降低,其准确性有所下降。当血小板计数≤15×10^9/L时,血液分析仪对血小板的计数有一定误差,应重视复核。有条件时建议用流式细胞术(FCM)复核。
出处 《临床检验杂志》 CAS CSCD 北大核心 2010年第2期151-152,共2页 Chinese Journal of Clinical Laboratory Science
  • 相关文献

参考文献3

  • 1ICSH. Cytometry panel platelet counting by RBC/platelet ratio method[J]. AJCP,2001,115(3) :460-464.
  • 2Harrison P, Ault KA, Chapman S, etal. An interlaboratory study of a candidate reference method for platelet counting[ J]. AJCP,2001, 115(3) :448-459.
  • 3朱忠勇.准确计数血小板方法学研究进展[J].国外医学(临床生物化学与检验学分册),2002,23(3):131-132. 被引量:134

二级参考文献8

  • 1Gill JE, Davis KA, Cowart WJ, et al. A rapid and accurate closed-tube immunoassay for platelets on an automated hematology analyzer[J]. Am J Clin Pathol, 2000,114: 47-56.
  • 2Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure[J]. Brit J Heamat, 2000,108 : 228-235.
  • 3Kunicka JE, Fischer G, Murphy J, et al. Improved platelet counting using two-dimensional laser light scatter[J]. Am J Clin Pathol, 2000,114 : 283-289.
  • 4England JM, Rowan RM, Bins M, et al. Recommended methods for the visual determination of white cell and platelet count[J]. Available at: 1988/WHO-LAB-88.3.
  • 5Hammerstrom J. Spurious platelet counts in acute leukemia with DIC due to cell fragmentation[J]. Clin Lab Heamat,1992,14: 239-243.
  • 6Matzdorff AC, Kuhnel G, Kemkes-Matthes B, et al.Quantitative assessment of platelets, platelet microparticles,and platelet aggregates with flow cytometry[J]. J Lab Clin Med,1998,131:507-517.
  • 7Dickerhoff R, von Ruecker A. Enumeration of platelets by multiparameter flow cytometry using platelet-specific antibodies and fluorescent reference particles [J]. Clin Lab Haemat, 1995,17:163-172.
  • 8Chapman DH, Hardin J, Miers M, et al. Reduction of the platelet review rate using the two-dimensional platelet method[J]. Am J Clin Pathol, 2001,115: 894-898.