三维与二维超声对睾丸体积测量准确性的对比研究 被引量:4

Accurate study of testicular volume measurement by three-dimensional and two-dimensional ultrasound
暂未订购
导出
摘要 目的评价三维超声、二维超声对睾丸体积测量的准确性。方法对20位拟行去势手术的晚期前列腺癌患者共40枚睾丸,术前使用二维超声对睾丸的长、宽、高(前后径)进行测量,使用椭球体公式计算睾丸体积;然后再使用三维超声自动测量程序测出睾丸容积。睾丸切除术后应用排水法测量睾丸的实际体积,并以此作为标准,评价三维超声和二维超声测量睾丸体积与睾丸实际体积的相关性及两种方法测量结果的准确性。结果两种超声测量方法所测睾丸体积与实际体积均高度相关;但二维超声测量的值较睾丸实际体积偏小,P<0.05,而三维超声容积探头所测体积与实际体积比较差异无统计学意义,P>0.05。结论三维超声对睾丸体积测量的准确性优于二维超声,所测数值能更好地代表实际睾丸体积;使用椭球体公式所计算的睾丸体积小于实际体积。 Objective To evaluate the accuracy of testicular volume measurement by three dimensional ultrasound and two dimensional ultrasound. Methods Forty testes in 20 patients with advanced stage prostate cancer were enrolled in this study. The length, width and height of testes were measured by two dimensional ultrasound, the testicular volume was calculated by round shape volume equation ( π/6 × lenght × width × height) , then the testicular volmne was measured by three dimensional ultrasound auto measurement program. The real volume of testis was measured by draining method as golden standard. The correlation between testicular volume measured by three dimensional and two dimensional ultrasound and the real volume was analyzed , and the accuracy was evaluated. Results Testicular volume measured by two different methods was closely correlated with the real volume, but the volume measured by two dimensional uhrasound was smaller than that of real volume, P 〈 0.05. While there was no significant difference between the volume measured by three dimensional ultrasound and draining method, P 〉 0.05. Conclusion The accuracy of testicular volume measurement by three dimensional ultrasound is better than that by two dimensional ultrasound,it is a better method to measure testicular volume. The volume obtained by equation of round shape volume is smaller than that of real testicle.
出处 《临床超声医学杂志》 2010年第2期92-94,共3页 Journal of Clinical Ultrasound in Medicine
关键词 睾丸 体积 超声检查 二维 三维 比较 Testicle Volume Uhrasonography, two dimension, three dimension Compariou
  • 相关文献

参考文献10

  • 1Hwang JY, Byun SS, Oh SJ, et al. Novel algorithm for improving accuracy of ultrasound measurement of residual urine volume according to bladder shape. Urology, 2004,64 (5) :887 - 891.
  • 2Hashimoto S, Goto H, Hirooka Y, et al. An evaluation of three - dimensional ultrasonography for the measurement of gallbladder volume. Am J Gastroenterol, 1999,94 ( 12 ) :3492 - 3496.
  • 3徐辉雄,徐作峰,吕明德,谢晓燕,刘丽,刘广健,黄蓓.三维超声容积自动测量技术的准确性及重复性研究[J].中国超声医学杂志,2003,19(3):232-236. 被引量:50
  • 4Martins WP, Ferriani RA, Barra DA, et al. Reliability and validity of tissue volume measurement by three -dimensional ultrasound: an experimental model. Ultrasound Obstet Gynecol, 2007, 29(2):210- 214.
  • 5Strommen K, Stormark TA, Iversen BM, et al. Volume estimation of small phantoms and rat kidneys using three - dimensional ultrasonography and a position sensor. Ultrasound Med Biol, 2004, 30 ( 9 ) : 1109 -1117.
  • 6Chang CH, Yu CH, Chang FM, et al, The assessment of normal fetal liver volume by three - dimensional ultrasound. Ultrasound Med Biol, 2003, 29(8) :1123 - 1129.
  • 7Ruano R, Martinovic J, Dommergues M, et al. Accuracy of fetal lung accessed by three - dimensional sonography. Ultrasound Obstet Gynecol, 2005, 26(7):725-730.
  • 8Oswald J, Schwenter C, Lunacek A, et al. Age and lean body weight related growth curves of kidneys using real - time 3 - dimensional ultrasound in pediatric urology. J Urol, 2004, 172(5) :1991 -1994.
  • 9Ying M, Sin MH, Pang SF. Sonographic measurement of thyroid gland volume: A comparison of 2D and 3D ultrasound. Radiography, 2005, 11 (4), 242 - 248.
  • 10童放,夏泽,夏炳兰,钱晓芹.不同旋转步长三维超声体积自动测量技术测量体积准确性的仿体实验研究[J].中国医学影像技术,2009,25(1):132-134. 被引量:9

二级参考文献20

  • 1吕小利,车岩,吴菊,锁成珍.三维彩色能量成像评价正常妊娠胎盘血流灌注[J].中国医学影像技术,2006,22(3):452-454. 被引量:17
  • 2Riecabona M, Nelson TR, Pretorid H, et al. Three-dimensional ultrasound : accuracy of distance and volume measurements, Ultrasound Obstet Gynecol, 1996,7(6):429-434.
  • 3Rankin RN, Fenster A, Downey DB, et al. Three-dimensional sonographic reconstruction: technichques and diagnostic applications. AJR Am J Roemgenol, 1993,161(4) :695-702.
  • 4Chang FM, Hsu KF, Ko HC, et al. Three-dimensional ultrasound assessments of fetal liver volume in normal pregnancy: a comparison of reproducibility with two-dimensional ultrasound and a search for a volume constant. Ultrasound Med Biol, 1997,23(3): 381-389.
  • 5Ruano R, Martinovic J, Dommergues M, et al. Accuracy of fetal lung volume assessed by three-dimensional sonography. Ultrasound Obstet Gynecol, 2005,26(7) :725-730.
  • 6[1]Liang RI, Chang FM, Yao BL, et al. Predicting birth weight by fetal upper-arm volume with use of three-dimensional ultrasonography. Am J Obstet Gynecol, 1997,177:632-638
  • 7[2]Hsieh YY, Chang CC, Lee CC, et al. Fetal renal volume assessment by three-dimensional ultrasonography. Am J Obstet Gynecol, 2000, 182: 377-379
  • 8[3]Riccabona M, Nelson TR, Pretorius DH, et al.Distance and volume measurement using three-dimensional ultrasonography. J Ultrasound Med, 1995,14:881-886
  • 9[4]Xu HX, Zhang QP,Lu MD, et al. Comparison of two-dimensional and three-dimensional sonography in evaluating fetal malformations. J Clin Ultrasound, 2002, 30:515-525
  • 10[5]Xu HX, Lu MD, Zhou YQ, et al. Three-dimensional gray scale volume rendering of the liver: preliminary clinical experience. J Ultrasound Med, 2002, 21:961-970

共引文献54

同被引文献38

引证文献4

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部