三种不同眼压计眼压测量值的比较 被引量:1

Comparison of intraocular pressure values measured by three different tonometers
暂未订购
导出
摘要 目的:比较分析Icare回弹式眼压计、GAT和DCT的眼压测量结果,探讨Icare回弹式眼压计的临床性能。方法:78例152眼分别用Icare,GAT,DCT3种眼压计进行眼压测量,然后根据测得的眼压高低分为高眼压、中眼压、低眼压3个组,对比分析3种眼压计的测量结果。结果:在全部受测者中Icare,GAT,DCT测得的眼压均值分别为19.16±5.03mmHg,18.41±4.52mmHg和17.23±3.69mmHg,每两种眼压计相比均有显著差别,但是彼此之间密切相关。高、中、低3个眼压组两种眼压计之间的差值均随着眼压的增高而增大。结论:使用Icare测量的眼压值准确可信,Icare,GAT和DCT的眼压值彼此之间具有良好的相关性。 AIM: To evaluate the clinical performance of Icare rebound tonometer by comparing the intraocular pressure (lOP) values mea-sured by Icare, Goldmann applanation tonometer (GAT) and dynamic contour tonometry (DCT). METHODS: IOP measurement was performed on 152 eyes of 78 subjects by Icare,GAT and DCT respectively. All subjects were devided into three groups ( high, medium and low IOP groups)according to their IOP values, the IOP measured by Icare,GAT and DCT were compared. RESULTS: The mean IOP of Icare,GAT and DCT were 19.16 5. 03mmHg, 18. 41 4. 52mmHg and 17. 23 ± 3.69mmHg respectively. Compared with each other,there were significant differences in lOP of Icare, GAT and DCT, however, there were closely correlations between each other. The mean lOP differences between each other increased along with the increase lOP in the high, medium and low lOP groups. CONCLUSION : The lOP measured by Icare was accurate and credible at the same time, there were closely correlations between the lOP of Icare and the lOP of GAT and DCT.
出处 《国际眼科杂志》 CAS 2010年第2期270-272,共3页 International Eye Science
关键词 眼压 眼压测量 ICARE回弹式眼压计 intraocular pressure intraocular pressure measurement Icare rebound tonometer
  • 相关文献

参考文献10

  • 1Harada H, Hirose N, Kubota T, et al. The influence of central corneal thickness and corneal curvature radius on the intraoeular pressure as measured by different tonometers: noncontact and goldmann applanation tonometers. J Glaucoma 2008 ; 17 ( 8 ) :619- 625.
  • 2A Kotecha ET, White JM, Shewry DF, et al. The relative effects of corneal thickniss and age on Glodmann applanation tonometer and dynamic contour tonometry. Br J Ophthalmol 2005 ; 89 : 1572-1575.
  • 3周和政,江文珊,张绍阳,晏颖,陈云辉,杜伟.ICARE回弹式眼压计与NCT眼压计测量结果的比较[J].国际眼科杂志,2009,9(3):484-485. 被引量:12
  • 4周和政,沈政伟,金中秋,王柏川,鄢俊杰.辩证地认识眼压在青光眼诊断中的地位[J].美中国际眼科杂志,2002,2(2):20-23. 被引量:13
  • 5Johannesson G, Hallberg P, Eklund A, et al . Pascal, ICare and Gold- mann applanation tonometrv-a comparative study. Acta Ophthalmol 2008 ;86( 6 ) :614-621.
  • 6Munkwitz S, Elkarmouty A, Hoffmann EM, et al . Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range. Graefes Arch Clin Exp Ophthalmol 2008 ; 246 ( 6 ) : 875 -879.
  • 7Pakrou N, Gray T, Mills R, et al . Clinical comparison of the Icare tonometer and Goldmann applanation tonometry.J Glaucoma 2008 ; 17 ( 1 ) : 43-47.
  • 8Queiros A, Gonzalez-Meijome JM, Fernandes P, et al. A comparison of central and peripheral intraocular pressure using rebound tonometry. Ophthal Physiol 2007 ;27(5 ) :506-511.
  • 9Schreiber W, Vorwerk CK,Langenbucher A, et al. A comparison of re- bound tonometer ( ICARE ) with tonometer and GAT. Ophthalmologe 2007 ; 104 ( 4 ) : 299-304.
  • 10Lpez-Caballero C, Contreras I, Muos-Negrete FJ, et al . Rebound tonometry in a clinical setting. Comparison with applanatlon tonometry. Esp Oftalmol 2007 ;82(5 ) :273-278.

二级参考文献13

  • 1高宗峰.安徽省桐城县青光眼流行病学调查[J].中华眼科杂志,1995,31(2):149-151. 被引量:26
  • 2胡铮,彭佑恩.北京市顺义县青光眼流行病学调查[J].中华眼科杂志,1989,25(2):115-119. 被引量:87
  • 3Schreiber W, Vorwerk CK, Langenbucher A, et al. A comparison of rebound tonometer(ICARE) with tonometer and GAT. Ophthalmologe 2007;104(4):299-304
  • 4Sahin A, Basmak H, Niyaz L, et al. Reproducibility and tolerability of the ICARE rebound tonometer in school children. J Glaucoma 2007;16(2):185-188
  • 5Toyohara K, Inoue K, et al. Comparison of ICARE rebound tonometer with GAT and non-contact tonometer. Atarashii (Journal of the Eye) 2007;24(3):355-359
  • 6López-Caballero C, Contreras I, Muos-Negrete FJ, et al. Rebound tonometry in a clinical setting.Comparison with applanation tonometry. Esp Oftalmol 2007;82(5):273-278
  • 7Queirós A, González-Méijome JM, Fernandes P, et al. A comparison of central and peripheral intraocular pressure using rebound tonometry. Ophthal Physiol Opt 2007;27(5):506-511
  • 8Davies LN, Bartlett H, Mallen EA, et al. Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 2006;84(2):206-209
  • 9傅伟才,张琳.开角型青光眼患者昼夜眼压监测及其意义[J].上海交通大学学报(医学版),2007,27(9):1163-1165. 被引量:3
  • 10王瑾,周和政,杨帆,曾波.PASCAL动态轮廓眼压计临床应用的初步报告[J].国际眼科杂志,2008,8(3):519-521. 被引量:8

共引文献20

同被引文献2

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部