左旋聚乳酸可吸收骨钉两种灭菌法效果评价

Evaluation of the sterilizing efficiency on absorbable poly-L-lactide screws with two sterilants
原文传递
导出
摘要 目的评价左旋聚乳酸可吸收骨钉经环氧乙烷灭菌和^60Co-γ辐照灭菌的效果。方法将左旋聚乳酸可吸收骨钉放人环氧乙烷灭菌器中灭菌,灭菌后将材料置于空气中静置5d,以清除残留的环氧乙烷;另将左旋聚乳酸可吸收骨钉采用^60Co-γ辐照灭菌装置灭菌,吸收剂量15kGy。然后对2种方法灭菌的骨钉分别进行性能表征。结果与左旋聚乳酸可吸收骨钉初始弯曲强度相比,经2种方法灭菌的骨钉弯曲强度均无明显下降(P〉0.05);与左旋聚乳酸可吸收骨钉初始分子量相比,经^60Co-γ辐照灭菌的骨钉分子量明显减小,下降约11%,表明材料发生降解;经环氧乙烷灭菌的骨钉分子量无明显变化,下降约3%,表明材料未发生明显降解。结论左旋聚乳酸可吸收骨钉以环氧乙烷灭菌为佳。 Objective To evaluate the sterilizing efficiency of absorbable poly-L-lactide (PLLA) screws sterilized with ethylene oxide and ^60Co-γ, radial. Methods One group of PLLA screws was sterilized in the ethylene oxide sterilization case. In order to eliminate the rudimental ethylene oxide, the sterilized screws was put in the air for five days. The other group of PLLA screws was sterilized with ^60Co-γ radial with absorbable dose of 15 kGy. Then all sterilized screws were characterized. Results Compared with the primal bend intensity of the PLLA screws, the bend intensity of sterilized screws did not descend remarkably (P〉0.05). Compared with the primal molecular weight of the PLLA screws, the molecular weight of screws with ^60Co-γ radiation sterilization remarkably decreased 11%. It indicated that the material had degraded. The molecular weight of screws with ethylene oxide sterilization only decreased 3%. It indicated that the material didn't degrade obviously. Conclusion Ethylene oxide sterilization is a better way for sterilizing absorbable PLEA screws.
出处 《国际生物医学工程杂志》 CAS 北大核心 2009年第4期201-205,共5页 International Journal of Biomedical Engineering
基金 基金项目:广州市科技计划项目(2006241C0671)
关键词 左旋聚乳酸可吸收骨钉 环氧乙烷 ^60Co-γ射线 灭菌 Absorbable poly-L-lactide screws Ethylene oxide ^60Co-γ radial Sterilization
  • 相关文献

参考文献18

  • 1Schnabelrauch M, Vogt S, Lareher Y, et al. Biodegradable polymer networks based on oligolactide macromers: synthesis, properties and biomedical applications[J]. Biomol Eng, 2002, 19(2-6): 295-298.
  • 2张红岗 肖芳.内固定器材是否取出的临床分析.中华实用医药杂志,2004,4(14):26-26.
  • 3Saikku-Backstrom A, Raiha JE, Valimaa T, et al. Repair of radial fractures in toy breed dogs with self-reinforced biodegradable bone plates, metal serews, and light-weight external coaptation[J]. Vet Surg, 2005, 34(1): 11-17.
  • 4Bergsma JE, de Bruijn WC, Rozema FR, et al. Late degradation tissue response to poly(L-lactide) bone plates and screws[J]. Biomaterials, 1995, 16(1): 25-31.
  • 5李二恪.生物材料在颌面部外伤治疗中的应用[J].口腔颌面外科杂志,1999,9(2):152-154. 被引量:7
  • 6Bos RRM, Rozema FR, Boering G, et al. Bone-plates and screws of bioabsorbable poly(L-lactide) an animal pilot study[J]. Br J Oral Maxillofac Surg, 1989, 27(6): 467-476.
  • 7Wang CH, Hsiue GH. New amphiphilic poly (2-ethyl-2-oxazoline)/poly (L-lactide) triblock copolymers [J]. Biomacromolecules, 2003, 4(6): 1487-1490.
  • 8Kim K, Yu M, Zong Xin-hua, et al. Control of degradation rate and hydrophilicity in electrospun non-woven poly(D,L-lactide) nanofiber scaffolds for biomedical applications[J]. Biomaterials, 2003, 24(27): 4977-4985.
  • 9Bostman OM. Absorbable implants for the fixation of fractures[J]. J Bone Joint Surg Am, 1991, 73(1): 148-153.
  • 10Suganuma J, Alexander H. Biological response of intramedullary bone to poly-L-lactic acid[J]. J Appl Biomat, 1993, 4(1): 13-27.

二级参考文献17

共引文献80

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部