期刊文献+

Effect of environment humidity to radon measurement with SSNTD 被引量:1

Effect of environment humidity to radon measurement with SSNTD
在线阅读 下载PDF
导出
摘要 For a passive radon detector of Solid State Nuclear Track Detector(SSNTD),the measurement results can be affected under an environment of high humidity.In order to correct this kind of measurement errors,experiments in normal radon laboratory were performed.The track films were etched in 6.25 mol/L NaOH solution at 70°C for 12 h,and rinsed with water for 6 h.The experiment results show that calibration factor of the SSNTD decreased with increasing humidity,because of increased stopping power of the moisture-absorbed film surface to αparticles.Therefore,the calibration factor K should be corrected from the calculated result. For a passive radon detector of Solid State Nuclear Track Detector (SSNTD), the measurement results can be affected under an environment of high humidity. In order to correct this kind of measurement errors, experiments in normal radon laboratory were performed. The track films were etched in 6.25 mol/L NaOH solution at 70℃ for 12 h, and rinsed with water for 6 h. The experiment results show that calibration factor of the SSNTD decreased with increasing humidity, because of increased stopping power of the moisture-absorbed film surface to a particles. Therefore, the calibration factor K should be corrected from the calculated result.
出处 《Nuclear Science and Techniques》 SCIE CAS CSCD 2009年第4期228-230,共3页 核技术(英文)
关键词 SSNTD 原子核物理学 能级 原子核 Solid state nuclear track detector, Humidity, Influence
  • 相关文献

参考文献8

  • 1Urban M, Piesch E. Radiat Prot Dosim, 1981, 1: 97-109.
  • 2Zhu R S. The theory and application of solid state nuclear track detector. Beijing: Science Press, 1987, 1:1-5 (in Chinese).
  • 3Li Y L, Li S Y. Radiat Prot Dosim, 1999, 5:377-381 (in Chinese).
  • 4Nan Q L, Zou S Y, Ding T X. Nucl Electron Detect Technol, 1999, 19: 72-74.
  • 5Wang X G, Luo Y S, Zhang H, et al. Nucl Sci Teclmol, 2005, 28: 319-323.
  • 6Fujimoto K. Radiat Prot Dosim, 1994, 55: 273-277.
  • 7Zhou X Y. Error theory and experiment data disposal. Beijing: Beijing aviation Institute Press, 1986, 69-70 (in Chinese).
  • 8Wu H S. The measuring method and application of radon. Beijing: Atomic Energy Press, 1995, 103 (in Chinese).