光激化学发光法检测甲胎蛋白实验性能评价 被引量:14

Analytical performance of light initiated chemiluminescence assay for serum α-fetoprotein
暂未订购
导出
摘要 目的评价光激化学发光法(LICA)检测血清甲胎蛋白(AFP)的实验性能。方法利用定标品和质控品对LICA检测AFP的实验性能进行评价;利用358份含不同浓度AFP的血清标本对LICA和罗氏电化学发光免疫法进行方法学对比评价。结果LICA检测AFP的分析敏感性为0.25μg/L,在平均浓度10和500μg/L的检测总变异系数(CV)均<5%。与参比方法所测AFP浓度差异无统计学意义,2种方法所测结果的对数值呈直线相关(r=0.966,P<0.001),对标本阴、阳性分类差异无统计学意义(P=0.824),有较好的一致性(Kappa=0.886,P<0.001)。结论LICA检测AFP具有良好的检测性能;与罗氏电化学发光法具有类似的实验性能。 Objective To evaluate the analytical performance of light initiated chemiluminescence assay (LICA) for serum α-fetoprotein (AFP). Methods The analytical performance was evaluated with control and calibrator materials, and methodological comparison was carried out with electrochemiluminescence immunoassay using serum samples from 358 subjects with various concentration of serum AFP. Results The analytical sensitivity was 0.25 μg/L with total coefficient variation (CV) of 〈 5% measured at approximately 10 and 500 μg/L. The two methods had no significant difference in AFP eoncentrations with significant linear correlation ( r = 0. 966, P 〈 0. 001 ) for log transformed values, and also had no significant difference in sample classification ( P = 0. 824) with high consistency (Kappa = 0. 886, P 〈 0. 001 ). Conclusions The LICA for AFP has very good analytical performance with diagnostic performance comparable with electrochemiluminescence immunoassay.
出处 《检验医学》 CAS 北大核心 2009年第8期578-581,共4页 Laboratory Medicine
关键词 光激化学发光法 甲胎蛋白 电化学发光免疫法 Light initiated chemiluminescence assay α-fetoprotein Electrochemilumineseence immunoassay
  • 相关文献

参考文献12

  • 1Ullman EF, Kirakossian H, Singh S, et al. Luminescent oxygen channeling immunoassay: measurement of particle binding kinetics by chemiluminescence [ J ]. Proe Natl Aead Sci, 1994, 91 (12) :3426-5430.
  • 2Ullman EF, Kirakossian H, Switchenko AC, et al. Luminescent oxygen channeling assay ( LOCITTM ) : sensitive, broadly applicable homogeneous immunoassay method[J]. Clin Chem, 1996, 42(9) :1518- 1526.
  • 3Dafforn A, Kirakossian H, Lao K. Miniaturization of the luminescent oxygen channeling immunoassay ( LOCI ^TM ) for use in multiplex array formats and other biochips [J]. Clin Chem, 2000, 46(9):1495-1497.
  • 4Poulsen F,Jensen KB. A luminescent oxygen channeling immunoassay for the determination of insulin in human plasma[ J]. J Biomol Screen, 2007, 12 (2): 240 -247.
  • 5Patel R, Pollner R, Keczer S, et al. Quantification of DNA using the luminescent oxygen channeling assay [J]. Clin Chem, 2000, 46(9) :1471-1477.
  • 6Beaudet L, Bedard J, Breton B, et al. Homogeneous assays for single-nucleotide polymorphism typing using AlphaScreen [ J ]. Genome Res, 2001, 11 (4) : 600-608.
  • 7Peppard J, Glickman F, He Y, et al. Development of a high-throughput screening assay for inhibitors of aggrecan cleavage using luminescent oxygen channeling ( AlphaScreen^TM ) [ J]. J Biomol Screen, 2003, 8(2) :149-156.
  • 8Wilson J, Rossi CP, Carboni S, et al. A homogeneous 384-well high-throughput binding assay for a TNF receptor using AlphaSereen technology [ J ]. J Biomol Screen, 2003,8 ( 5 ) :522-532.
  • 9马宏伟,赵卫国,潘柏申.血清肌红蛋白光激化学发光免疫测定法的建立[J].检验医学,2006,21(1):55-57. 被引量:11
  • 10马宏伟,赵卫国,潘柏申.血清心肌肌钙蛋白Ⅰ光激化学发光免疫测定法的建立[J].检验医学,2007,22(4):398-401. 被引量:7

二级参考文献20

  • 1马宏伟,赵卫国,潘柏申.亲和层析法分离纯化人心肌肌钙蛋白Ⅰ[J].检验医学,2007,22(3):304-307. 被引量:3
  • 2Panteghini M, Pagani F, Bonetti G. The sensitivity of cardiac markers: an evidence-based approach[J].Clin Chem Lab Med, 1999, 37 : 1097-1106.
  • 3Ullman EF, Kirakossian H, Singh S, et al. Luminescent oxygen channeling immunoassay, measurement of particle binding kinetics by chemiluminescence[J].Proc Natl Acad Sci, 1994, 91 :5426-5430.
  • 4Zaninotto M, Pagani F, Altinier S, et al. Multicenter evaluation of five assays for myoglobin determination[J]. Clin Chem, 2000,46:1631-1637.
  • 5Wong SS. Strategic utilization of cardiac markers for the diagnosis of acute myocardial infarction[J]. Ann Clin Lab Sci, 1996, 26: 301-312.
  • 6Le Moigne F, Beauvieux MC, Derache P, et al. Determination of myoglobin : comparative evaluation of the new automated VIDAS assay with two other immunoassays[J]. Clin Biochem, 2002,35:255-262.
  • 7Panteghini M, Linsinger T,Wu AH, et al. Standardization of immunoassays for measurement of myoglobin in serum. Phase Ⅰ: evaluation of candidate secondary reference materials[J]. Clin Chim Acta, 2004,341:65-72.
  • 8Brunori M, Bourgeois D, Vallone B. The structural dynamics of myoglobin[J]. J Struct Biol, 2004,147 : 223-234.
  • 9Takeda S,Yamashita A,Maeda K,et al.Structure of the core domain of human cardiac troponin in the Ca(2+)-saturated form[J].Nature,2003,424(6944):35-41.
  • 10Apple FS,Murakarni MM.The diagnostic utility of cardiac biomarkers in detecting myocardial infarction[J].Clin Comerstone,2005,7(1):25-30.

共引文献15

同被引文献107

引证文献14

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部