期刊文献+

高含量金样品均匀性的分析检验 被引量:7

Homogeneity Test for Samples with High-content of Gold
在线阅读 下载PDF
导出
摘要 利用王水溶样,残渣经氢氟酸、高氯酸分解-活性炭吸附-原子吸收光谱法测定了高含量金样品“哈测金-1”的15个子样中的全金量,并对子样进行了平行双份分析。考察了天平称量精度、玻璃器皿刻度精度和活性炭吸附率造成的测量不确定度,并估算出合成不确定度范围为2.5%~4.2%;论证了方法的可行性。对单个子样重复分析结果的可靠性及其算术平均值的极限误差进行了评估,计算出样品均匀性质量参数为0.70,依此判定该样品的均匀性质量等级为“良好”。 15 samples, named "HACEAu-1", with high-content of gold were digested with aqua regia and the residues were decomposed with HF-HC104. Gold in the sample solution was preconcentrated by activated carbon adsorption and determined by FAAS. Duplicate analysis for 15 samples was carried out and the measurement uncertainty for the analytical results was evaluated. The main sources of measurement uncertainty came from sub-uncertainties of the analytical balance, scale accuracy of the analytical glassware, gold adsorption rate by activated carbon. The composed uncertainty was calculated with a range of 2.5% to 4.2%, which means that the method was feasible. The reliability for the results from duplicate analysis of single sample and the limiting error of its arithmetic mean value were also evaluated. The homogeneity quality coefficient of the sample was calculated as 0.70, which means that the sample was in "good" homogeneity quality grade.
出处 《岩矿测试》 CAS CSCD 北大核心 2009年第4期388-390,共3页 Rock and Mineral Analysis
基金 中国地质调查局科技攻关项目资助(1212010660902-01)
关键词 高含量金 哈测金-1 均匀性 high-content of gold HACEAu-1 homogeneity
  • 相关文献

参考文献6

二级参考文献5

  • 1陈青川.活性炭对碘化亚金离子的富集[J].黄金地质科技,1993(3):68-70. 被引量:3
  • 2招远金矿化验室.活性炭吸附柱分离和碘量法测定金[J].分析化学,1976,4(6):445-446.
  • 3中国有色金属工业总公司矿产地质研究院.MGI-Au-01-MGI-Au-04金银地质标准参考样品的研制.黄金,1985,5(4):54-65.
  • 4熊昭舂.地质试料中痕量—超痕量金分析的新进展[J].地质实验测试通讯,(2):6-6.
  • 5郑大中.活性碳吸附金机理的探讨[J].地质地球化学,1991(5):59-61. 被引量:12

共引文献23

同被引文献106

引证文献7

二级引证文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部