期刊文献+

缝管捕集原子吸收光谱法测定环境样品中微量Pb、Cu和Cd 被引量:12

DETERMINATION OF TRACE Pb,Cu and Cd IN ENVIRONMENTAL SAMPLE BY SLOTTEDTUBE ATOMTRAPPING ATOMIC ABSORPTION SPECTROMETRY
在线阅读 下载PDF
导出
摘要 研究了缝管原子捕集原子吸收光谱法测量Pb、Cu和Cd的最佳条件。结果表明,在选定的条件下,捕集1min,测得Pb、Cu和Cd的特征浓度分别为182×10-3、126×10-3和67×10-5mg/L,比常规火焰原子吸收光谱法的灵敏度分别提高110、39和150倍,用本方法测定地表水、茶叶和人发等环境样品中的微量Pb、Cu和Cd,结果满意。 The optimal measuring conditions of slottedtube atomtrapping atomic absorption spectrometry of Pb,Cu and Cd were examined in detail.The results show that the characteristic concentrations measured after trapping for 1min under definite conditions were 182×10-3,126×10-3 and  67×10-5mg/L for Pb,Cu and Cd,respectively.Thus slotted tube atomtrapping AAS yields increase in sensitivety for Pb,Cu and Cd by 110,39 and 150 fold,respectively,in comparison with the conventional FAAS.Trace elements Pb,Cu and Cd in earth surface water,tea and hair were determined by slotted tube atomtrapping.Satisfactory results were obtained.
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 1998年第3期347-350,共4页 Spectroscopy and Spectral Analysis
基金 湖北省自然科学基金
关键词 原子吸收光谱 测定 分析 环境样品 Atomtrapping, Atomic absorption spectrometry, Pb, Cu, Cd
  • 相关文献

参考文献5

二级参考文献10

  • 1黄淦泉,分析化学,1986年,14卷,198页
  • 2黄淦泉,分析化学,1986年,14卷,593页
  • 3李银玉,分析化学,1986年,16卷,1006页
  • 4刘永铭,分析化学,1986年,14卷,55页
  • 5黄淦泉,光谱学与光谱分析,1987年,7卷,60页
  • 6李银玉,分析化学,1988年,16卷,11期,1006页
  • 7刘永铬,分析化学,1988年,16卷,1期,55页
  • 8黄淦泉,光谱学与光谱分析,1987年,7卷,60页
  • 9黄淦泉,分析化学,1986年,14卷,8期,593页
  • 10黄淦泉,李云清.开缝石英管原子化技术的改进[J].高等学校化学学报,1992,13(8):1057-1059. 被引量:7

共引文献13

同被引文献1708

引证文献12

二级引证文献33