摘要 目的:本文研究了我院X-刀治疗颅内肿瘤过程中CT定位的精度及CT扫描方法。方法:应用人体头颅模型内特定标记物测定CT定位的精度,在临床工作中应用不同的扫描方法对颅内肿瘤进行CT定位,观察其效果。结果:利用头颅模型测定的BRW头环CT定位精度为0.65mm,最大误差为1.09mm。对35例颅内肿瘤病人进行X-刀CT定位,选择了直接强化、延迟3分钟和5分钟强化的方法,分别扫描脑膜瘤、听神经瘤、转移瘤、胶质瘤和垂体瘤,多数患者肿瘤强化明显,肿瘤边界显示清晰,效果满意。结论:我院X-刀治疗CT定位精度高,对颅内肿瘤患者的肿瘤定位效果较好,符合X-刀定位的要求。 Purpose:This article study the accuracy and method of CT scanning in localization of intracranial tumor in X knife treatment.Method:Measure the CT localization acuracy using special markers in skullphantom and clinical use of different CT scanning method inlocalization of intracranial tumor during X knife treatment.Results:The accuracy of CT localization of BRW headring using skull phantom is 065mm averagely,maximum error is 109mm During X knife treatment,the method of CT scanning for localization of intracranial tumor is:immediat enhancement for meningiomas,acoustic neurofibromas and metastatic tumors,delayed 3 minutes enhancement for gliomas and delayed 5 minutes enhancement for pituitary tumors.Most of tumurs of 35 patients have been obviously enhanced,and show clear rims,all obtain satisfactory results for X knife treating.Conclusion:The accuracy and results of CT localization of intracranial tumor is suitable and dependentable for X knife treatment in our hospital.
出处 《中国医学影像技术》 CSCD 北大核心 1998年第3期172-174,共3页 Chinese Journal of Medical Imaging Technology
关键词 CT定位 X刀 颅内肿瘤 精度 治疗 Localization\ \ X knife\ \ Intracranial tumor\ \ Accuracy
  • 相关文献

同被引文献52

  • 1王云彦,张庆林,江玉泉,李刚.X-刀二次等中心系统的验证[J].中国微侵袭神经外科杂志,1998,3(3):175-176. 被引量:2
  • 2赵奇煌,陈炳桓,杨炯达,阎润民,梁继军,王俊岭,肖泽久,陈亚林,钱渡江,陈滨.颅内深部中线及中线旁病变的X-刀分次治疗的临床研究[J].首都医学院学报,1995,16(2):119-123. 被引量:3
  • 3曹丹庆 蔡祖龙.全身CT诊断学(第一版)[M].北京人民军医出版社,1996.142-144.
  • 4吴恩惠.头部CT诊断学(第二版)[M].北京:人民卫生出版社,1995.40-168.
  • 5Leksell L.The stereotactic method and radiosurgery of the brain. Aeta Chit Seand, 1951,102: 316-322.
  • 6Winston KR., Lutz W. Linear accelerator as a neurosulgieal tool for stereotaetic radiosurgery. Neurosurgey, 1988,22: 254-259.
  • 7Maeiumas R, Galloway RL, latimer J. The application accuracy of stereotactic frames. Neurosurgery, 1994,35: 682-694.
  • 8Hartmann GH, Baucer-Kirpes B, Serago CF, et al. Precision and accuracy of stereotactic convergent beam irridiation from a linear accelerator. Int J Radiat Oncol Biol Phys, 1993,28: 481-492.
  • 9Podgorsak EB, Pike GB, Pla M, et al. Radiosurgery with photon beams:physical aspects and accuracy of linear accelerators. Radiother Oncol,1990,17: 349-358.
  • 10Lutz W, Winston KR, Maleki N. A system for stereotacfic radiosurgery with a linear accelerator. Int J Radiat Oncol Biol Phys, 1988,14: 373-381.