折射/衍射球柱有晶体眼人工晶体的设计 被引量:3

Design of refractive/diffractive spherocylindrical phakic intraocular lens
在线阅读 下载PDF
导出
摘要 为了提高白光下人眼的视觉质量,设计了一款折射/衍射球柱有晶体眼人工晶体(PIOL)。根据实际人眼的角膜地形图、眼内各组分的轴向间距和波前像差,运用Zemax光学设计软件,构建了个性化眼模型;然后,运用该模型,设计了球柱PIOL和折射/衍射球柱PIOL。研究表明:与球柱PIOL相比,折射/衍射球柱PIOL能够全面提高人眼的白光调制传递函数(MTF)和非中心波长下的MTF,有利于获得良好的视觉质量。研究还表明:折射/衍射球柱PIOL只需要在4 mm区域设计二元面就能够满足要求。最后,分析了加工工艺问题。折射/衍射球柱PIOL在先进视觉矫正方面具有潜在的应用价值。 A refractive/diffractive spherocylindrical Phakic Intraocular Lens (PIOL) is designed to improve the white light visual quality of the human eye. On the basis of measured individual corneal data, eyeball depths and wavefront aberrations, an individual eye model is constructed with the optical design software Zemax,then the spherocylindrical PIOL and the refractive/diffractive spherocylindrical PIOL are designed by the established eye model. A comparison with spherocylindrical PIOL shows that the refractive/diffractive spherocylindrical PIOL can optimize overall white light Modulation Transfer Functions (MTF) and the MTFs of non-center wavelengths to improve the visual quality. The experiment also shows that a zone of 4 mm is enough for the binary structure. Finally, the fabrication issues are discussed. Obtained results indicate that the designed refractive/diffractive spherocylindrical PIOL has great potential ities in advanced vision correction.
出处 《光学精密工程》 EI CAS CSCD 北大核心 2009年第4期701-706,共6页 Optics and Precision Engineering
基金 国家自然科学基金资助项目(No.60438030,No.60777011) 高等学校博士学科点专项科研基金资助项目(No.20050055002)
关键词 视觉光学 色差 个性化眼模型 调制传递函数 有晶体眼人工晶体 visual optics chromatic aberration individual eye model Modulation Transfer Function (MTF) Phakie Intraocular Lens (PIOL)
  • 相关文献

参考文献10

  • 1MORRIS G M, NORDAN T L. Phakic Intraocular Lens: the new focus in refractive surgery [J].Optics and Photonics News, 2004, 15(9) :26-31.
  • 2MCLELLAN J S, MARCOS S, PRIETO P M, et al.. Imperfect optics may be the eye's defense against chromatic blur[J].Nature, 2002, 417 (5):174-176.
  • 3NORBERTO L, ROBERT M. New intraoeular lens for achromatizing the human eye[J].J.Cataract Refract. Surg., 2007,33(3):1296-1302.
  • 4NAVARRO R, SANTAMARIA J, BESCOIS J,Accommodation-dependent model of the human eye with aspherics[J]. J. Opt. Soc. Am. A, 1985, 2 (8) :1273-1281.
  • 5GUO H Q, WANG ZH Q, ZHAOQ L, et al.. In-dividual eye model based on wavefront aberration[J].Optik,2005,116(1) :80-85.
  • 6WANG Y, WANG ZH Q, WANG Y, et al.. Intraocular lens design for treating high myopia based on individual eye model[J]. Optik, 2007,118 ( 1 ) : 88-93.
  • 7HOLLADAY J T, PIERS P A, KORANYI G. A new intraocular lens design to reduce spherical aber ration of pseudophakie eye[J].J. Cataract Re fract. Surg., 2002,18(6):683-691.
  • 8郭欢庆,王肇圻,赵秋铃,全薇,王雁.基于主观式测量人眼波前像差的眼模型研究[J].光子学报,2005,34(11):1666-1669. 被引量:15
  • 9王肇圻,张轶楠,傅汝廉,孙强.折/衍混合Petzval光电摄像物镜设计[J].光学精密工程,2005,13(1):1-4. 被引量:22
  • 10WANG ZH Q, ZHANG H J, FU R L, et al..Hybrid diffractive-refractive ultra-wide-angle eyepieces[J]. Optik, 2002,113(1):159-162.

二级参考文献21

  • 1全薇,王肇圻,宋贵才,凌宁,傅汝廉.用SVGA1薄膜晶体管液晶显示器矫正人眼波像差[J].光子学报,2004,33(12):1445-1448. 被引量:12
  • 2岳宏,戴士杰,崔庆华,史建儒,王仲民.光学小波变换在视觉系统的应用研究[J].光子学报,2005,34(3):455-459. 被引量:3
  • 3Pantazis M. Visual instrumentation. New York: McGraw-Hill, 1999.106~110
  • 4Blaker J W. Toward an adaptive model of the human eye.J Opt Soc Am, 1980,70(2): 220~223
  • 5Lotmar W. Theoretical eye model with aspherics.J Opt Soc Am, 1971, 61(11): 1522~1529
  • 6Kooijman A C. Light distribution on the retina of a wide-angle theoretical eye.J Opt Soc Am, 1983, 73(11): 1544~1550
  • 7Navarro R,Santamaria J,Bescos J. Accommodation-dependent model of the human eye with aspherics.J Opt Soc Am A, 1985, 2(8):1273~1281
  • 8Hwey-Lan Liou,Brennan N A. Anatomically accurate, finite model eye for optical modeling.J Opt Soc Am A, 1997, 14(8): 1684~1695
  • 9Junzhong Liang, Bernhard Grimm, Stefan Goelz,et al.Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor.J Opt Soc Am A, 1994,11(7): 1949~1957
  • 10Prieto P M, Fernando Vargas-Martin, Stefan Goelz,et al.Analysis of the performance of the Hartmann-Shack sensor in the human eye.J Opt Soc Am A, 2000, 17(8): 1388~1398

共引文献32

同被引文献24

  • 1王肇圻,张轶楠,傅汝廉,孙强.折/衍混合Petzval光电摄像物镜设计[J].光学精密工程,2005,13(1):1-4. 被引量:22
  • 2翟琴华,罗鑫,谢兵.LASIK矫治PRK术后屈光度欠矫及回退[J].眼外伤职业眼病杂志,2005,27(4):284-286. 被引量:1
  • 3赵顺龙,王肇圻.超轻小型投影式头盔显示系统折-衍混合物镜设计[J].光学学报,2006,26(2):249-253. 被引量:12
  • 4范长江,王肇圻,吴环宝,张梅.红外双波段双层谐衍射光学系统设计[J].光学学报,2007,27(7):1266-1270. 被引量:28
  • 5BEDFORD R E, WYSZECKI G. Axial chromatic aberration of the human eye [J].J. Opt. Soc. Am. 1957,47(6):564-565.
  • 6THIBOS L N, BRADLEY A, STILL D L, et al Theory and measurement of ocular chromatic aberration [J]. Vision Research, 1990,30(1) :33-49.
  • 7RYNDERS M, LIDKEA B, CHISHOLM W, et al.. Statistical distribution of foveal transverse chromatic aberration, pupil centration, and angle ψ in a population of young adult eyes [J]. J. Opt. Soc. Am. A, 1995,12(10) :2348-2357.
  • 8THIBOS L N, YE M, ZHANG X X, et al.. The chromatic eye: a new reduced-eye model of ocular chromatic aberration in humans[J]. Applied Optics, 1992,31(19) :3594-3600.
  • 9MEETEREN A V, DUNNEWOLD C J W. Image quality of the human eye for eccentric entrance pupils [J]. Vision Research. , 1983,23(5) :573-579.
  • 10OGBOSO Y U, BEDELL H E. Magnitude of lateral chromatic aberration across the retina of the human eye [J]. J. Opt. Soc. Am. A,1987,4(8):1666-1672.