期刊文献+

仪器对凝血酶原时间国际标准化比值的影响及校正 被引量:12

The instrument's influence and rectifying on the prothrombin time of international normalised ratios
暂未订购
导出
摘要 为了评价血凝仪对国际标准化比值(INR)的影响,模拟WHO标定凝血质的参比方法,用一种已知国际敏感指数(ISI)的凝血质对两种血凝仪的特异性ISI进行标定,并以INR形式报告PT结果。结果表明,未经标定前,两种血凝仪测得结果有显著性差异(P<0.05),标定其系统的ISI后,再以公式INR=PRISI得出其结果,结果无差异。说明不同血凝仪对INR有一定影响。 In order to evaluate the coagulometers'influence on INR. by lmitating the reference method of World Health Organization(WHO) demarcation thromboplastin, has used an already known thromboplastin of the international Sensitivity index(ISI) to demarcate the distinctine quality ISI of two kinds of coagulometers, and has made the report of PT result in the form of INR. It shows thst there lies a marked difference between the two kinds of coagulometers before being demarcated. After demarcating the ISI of their systems and getting the result in the formula as INR=PR ISI , the difference disappears. This result shows that different coagulometers have some influence on the INR, and the demarcation of the instruments can avoid the influence partially.
出处 《临床检验杂志》 CAS CSCD 北大核心 1998年第2期89-90,共2页 Chinese Journal of Clinical Laboratory Science
关键词 国际标准化比值 国际敏感指数 血凝仪 International Normalised Ratios International Sensitivity Index Coagulometers
  • 相关文献

同被引文献45

引证文献12

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部