摘要
目的验证批量监测系统能否取代包内指示物。方法自2007年4月17日-2008年7月29日,对消毒供应室7台灭菌器采用过程监测系统对每一批次的灭菌进行效果监测,同时采用包内指示物做对比验证。结果历经15个月,批量监测7057批次,合格7051批次,合格灭菌包166 870包,包内指示物均合格;其中敷料包56 600包,器械包110 270包,合格率达100%;经对比验证,批量过程监测与包内指示物监测均达成一致。结论通过大批量的试验对比数据证实了批量过程监测系统的效果可靠,可取代包内指示物。
OBJECTIVE To verify the effect of batch monitoring system insteod of the indicator in the package. METHODS From Apr 17, 2007 to Jul 29, 2008, the batch monitoring system was used to monitor the sterilization effect of each batch, at the same time, inside each package we placed the indicator as control. RESULTS A total of 7057 batches were monitored, in which 7051 were eligible, and among 166870 packages in all, the indicators were all eligible, which included 56600 dressing packages and 110270 instrument packages. There was no difference between batch monitoring system and indicators. CONCLUSIONS The effect of batch monitoring system is credible, and the indicator inside package can be replaced.
出处
《中华医院感染学杂志》
CAS
CSCD
北大核心
2009年第6期655-656,共2页
Chinese Journal of Nosocomiology
关键词
批量监测
灭菌过程
效果验证
Batch monitoring
Sterilization process
Effect validation