主动固定螺旋电极植入术中的阈值变化 被引量:9

Change of the thresholds of helix electrodes during the implantation
暂未订购
导出
摘要 目的观察主动固定螺旋电极的起搏阈值在起搏器植入术中的变化规律。方法72例共植入74根螺旋电极。电极到位后,分别测试螺旋旋出前后的起搏阈值。结果螺旋旋出前的起搏阈值为0.56±0.30V,旋出后即刻为0.97±0.49V,旋出后5,10min分别为0.46±0.29V和0.36±0.20V。与旋出前相比,旋出后即刻的阈值明显增高(P〈0.01)。随后阈值逐渐下降,在旋出后5min可基本恢复至旋出前水平。同时右室不同部位的起搏参数无明显差异。结论主动固定螺旋电极在螺旋旋出前后阈值变化明显,5~10min后达稳定状态。 Objective To investigate the threshold variation of the helix electrodes during the implantation. Methods Seventy-two patients with 74 helix electrodes were enrolled in our study. After the cardiac pacing leads were put on the suitable sites of the ventricle or the atrium, the thresholds were tested when the electrode leads were screwed in before and after 0,5, and 10 minutes. Results The thresholds of the helix electrodes were: 0.56 ± 0.30 V before screwed, 0.97 ±0.49 V, 0.46 ± 0.29 V and 0.36 ± 0.20 V after screwed immediately , 5 and 10 minutes after screwed later. Conclusion The thresholds of helix electrodes increased significantly after being screwed immediately, but decreased gradually to the original level before being screwed after 5-10 minutes.
出处 《中国心脏起搏与心电生理杂志》 2008年第6期499-500,共2页 Chinese Journal of Cardiac Pacing and Electrophysiology
关键词 心血管病学 心脏起搏 主动固定电极 起搏阈值 Cardiology Cardiac pacing Helix electrode Pacing threshold
  • 相关文献

参考文献5

二级参考文献37

  • 1郭诗东,华伟,张澍,王方正,陈柯萍,吴国玫,陈新.右心室间隔部起搏的血流动力学研究[J].中国介入心脏病学杂志,2005,13(2):81-83. 被引量:37
  • 2汪芳,张建军,金炜,Jackson K,Lieberman R,孙宝贵.采用主动固定电极导线行右室流出道间隔部起搏的临床应用[J].中国心脏起搏与心电生理杂志,2006,20(2):151-152. 被引量:42
  • 3Cameron J,Mond H,Ciddor G,et al. Stiffness of the distal tip of bipolar pacing leads. PACE, 1990,13 : 1915-1920.
  • 4Mond HG,Hua W,Wang CC. Atrial pacing leads:the clinical contribution of steroid elution. PACE,1995 ,18 :1601-1608.
  • 5Hua W, Mond H, Sparks P. The clinical performance of three designs of atrial pacing leads from a single manufacturer: the value of steroid elution. Eur JCPE,1996,6:99-103.
  • 6Mond H,Stokes KB. The steroid-eluting electrode: a 10-years experience. PACE, 1997,19:1016-1020.
  • 7Gillis AM, Rothschild JM, Hillier K,et al. A randomized comparison of a bipolar steroid-eluting electrode and a bipolar microporous platinum electrode. PACE,1993,16 : 964-970.
  • 8Wish M,Swartz J,Cohen A,et al. Steroid tipped leads versus porous platinum permanent pacemaker leads: a controlled study. PACE, 1990,13 : 1887-1890.
  • 9Mond H, Stokes K, Helland J, et al. The porous titanium steroid eluting electrode: a double blind study assessing the stimulation threshold effects of steroid. PACE, 1988,11 : 214-219.
  • 10Sutton R,Guneri S. The impact of steroid eluting leads on long term pacing in the atrium and ventricle. Eur JCPE,1991,1:10-15.

共引文献52

同被引文献57

引证文献9

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部