摘要
采用原子荧光光谱法测定水处理剂聚氯化铝中砷。实验结果表明,线性关系良好(在标准曲线0.0~30.0μg/L 内,平均相关系数γ为0.9996),精密度较好(相对标准偏差为1.76%~2.66%),准确度、灵敏度较高(加标回收率为103.5%~107.7%,最低检出限为0.08μg/L),是测定水处理剂聚氯化铝中砷的较好方法。
With atomic fluorescence spectrometetic determination of Arsenic in Water treatment chemical- Poly aluminium chloride was carried out,The results obtained showed that linearity is well (the average correlate is 0.9996 in standard cure 0.0-30.0μg/L).The precision is better(the relative standard deflectivity is 1.76%-2.66%).The accuracy and sensitivity are higher (the recovery is 103.5%-107.7%,detection limit is 0.08μ g/L).It is better method for the determination of Arsenic in Water treatment chemical-poly aluminium chloride.
出处
《城镇供水》
2008年第7期63-65,共3页
CITY AND TOWN WATER SUPPLY
关键词
原子荧光光谱法
砷的测定
水处理剂聚氯化铝
氢化物发生
atomic fluorescence spectrometry
determination of Arsenic
Water treatment chemical- Poly aluminium chloride
hydride generation