期刊文献+

后循环缺血的MR电影相位对比法血流定量研究 被引量:5

Study of Flow Quantification in Posterior Circulation Ischemia:Measurement with Cine Phase-Contrast MR Imaging
暂未订购
导出
摘要 目的:探讨MR电影相位对比法测量后循环缺血患者双侧椎动脉血流流速、血流量等血流动力学参数的准确性及后循环缺血的原因。方法:28例经临床诊断为后循环缺血的患者及5例同年龄组健康志愿者,使用头部线圈和周围脉搏门控技术,采用MR电影相位对比法测量双侧椎动脉的血流动力学参数,并与经颅多普勒(TCDU)检查,结果进行比较,同时分析椎动脉血流量与血管狭窄程度的相关性。结果:MR电影相位对比和TCDU两种技术所测量的双侧椎动脉流速呈高度相关(左侧r=0.887,P=0.013;右侧r=0.785,P=0.027)。MR电影相位对比法测得病变组左右椎动脉平均流速分别为(16.16±7.99)和(15.72±9.92)cm/s,平均血流量分别为(66.63±45.23)和(68.57±38.84)ml/min。MR电影相位对比法所测血流量正常对照组与后循环缺血组之间差异有统计学意义。椎动脉血流量与血管狭窄程度无明显相关性(r=-0.144,P=0.654)。结论:使用MR电影相位对比法可无创性准确评价后循环缺血患者血流动力学改变,后循环的平均椎动脉血流量与血管狭窄程度低度相关。 Objective,. To evaluate the accuracy of measuring the hemodynamic parameters including flow velocity and flow volume of bilateral vertebral arteries in patients with posterior circulation isehemia (PCI) by using cine phase-contrast imaging and to study the causes of PCI. Methods: Cine phase-contrast MRI and trans-cranial doppler ultrasonography (TCDU) were performed in 28 patients with clinical diagnosed PCI and 5 healthy volunteers of the same age. Head coil and periphery pulse gate control technology were used. The results of hemodynamic parameters measured by both techniques were compared. The flow volume were correlated with the degree of stenosis of vertebral arteries. Results:In pauents with PCI, flow velocity of bilateral vertebral artery assessed by using Cine phase-contrast MRI and TCDU ultrasonography showed good correlation (left vertebral artery (LVA), r=0. 887. P =0. 013 ; Right vertebral artery (RVA) , r=0. 785. P=O. 027). The mean flow velocity measured by cine phase-contrast MRI were as follows: LVA (16. 16±7. 99) cm/s; RVA (15.72±9.92) cm/s;blood flow:LVA (66. 63±45.23) ml/min,RVA (68.57+38.84) ml/min. Significant statistical difference existed in the blood flow volume measured by eine phase-contrast MRI between PCI patients and volunteers No significant correlation was assessed between the flow volume and the degree of stenosis of vertebral artery (r=-0. 144, P=0. 654). Conclusion:Measurement of hemodynamic parameters could be obtained by using cine phase-contrast MR scan- ning,which is an non-invasive technique, and hemodynamic changes in patients with PCt could be accurately evaluated. However,blood flow and vascular stenosis was poorly correlated.
出处 《放射学实践》 2008年第11期1265-1268,共4页 Radiologic Practice
关键词 磁共振成像 电影相位对比法 血流动力学 超声检查 多谱勒 经颅 Magnetic resonance imaging Cine phase-contrast Hemodynamics Ultra sonography, doppler, transcranial
  • 相关文献

参考文献16

  • 1李焰生.“椎基底动脉供血不足”——这一概念还有用吗?[J].中华脑血管病论坛,2005,3(2):137-146. 被引量:4
  • 2Soustiel JF, Glenn TC, Vespa P, et al. Assessment of Cerebral Blood Flow by Means of Blood-Flow-Volume Measurement in the Internal Carotid Artery: Comparative Study With a 133 Xenon Clearance Technique[J]. Stroke, 2003,34(8) : 1876-1880.
  • 3Stella Sin Yee Ho, Yu Leung Chan, David Ka Wai Yeung, et al. Blood Flow Volume Quantification of Cerebral Ischemia: Comparison of Three Noninvasive Imaging Techniques of Carotid and Vertebral Arteries[J]. AJR, 2002,178(3): 551-556.
  • 4Zierler BK, Kirkman TR, Kraiss LW, et al. Accuracy of Duplex Scaning for Measurement of Arterial Volume Flow [J]. Vasc Surg, 1992,16(4): 520-526.
  • 5Oktar SO,Yticel C,Karaosmanoglu D. Blood-Flow Volume Quantification in Internal Carotid and Vertebral Arteries: Comparison of 3 Different Ultrasound Techniques with Phase-Contrast MR Imaging[J]. AJNR,2006,27(2):363-369.
  • 6朱晓黎,沈天真,陈星荣.MR脑血流定量在老年人无症状性脑缺血中的应用研究[J].临床放射学杂志,2005,24(9):759-764. 被引量:2
  • 7Michiel W,Jos M,Alphons J. Phase-contrast Magnetic Resonance Flow Quantification in Renal Arteries[J]. Hypertension, 2003,41 (1) : 114-118.
  • 8郑建明,史树贵,迟路湘,范文辉.105例椎基底动脉供血不足的脑血管造影分析[J].卒中与神经疾病,2006,13(3):172-174. 被引量:3
  • 9Gita A. Varaprasathan,Philip A,et al. Quantification of Flow Dynamics in Congenital Heart Disease: Applications of Velocity-encoded Cine MR[J]. RadioGraphics, 2002,22 (4) : 895 - 905.
  • 10Philipp B, Hermann K, Jurgen G, et al. Rapid Left to-Right Shunt Quantification in Children by Phase-Contrast Magnetic Resonance Imaging Combined With Sensitivity Encoding (SENSE) [J]. Circulation,2003,108(11) :1355-1361.

二级参考文献19

  • 1陈鸿儒,查振刚,董炘.椎动脉型颈椎病诊治研究进展[J].临床骨科杂志,2004,7(3):344-347. 被引量:11
  • 2夏军,周义成.影像学在眩晕诊断中的应用[J].卒中与神经疾病,2004,11(5):313-315. 被引量:5
  • 3何济华.无症状脑梗塞[J].国外医学(脑血管疾病分册),1995,3(4):206-209. 被引量:13
  • 4Plum F. What causes infarction in ischemic brain? Neurology, 1983, 33:222
  • 5Krejza J, Mariak Z, Walecki J,et al. Transcranial colour doppler sonography of basal cerebral arteries in 182 healthy subjects: age and sex variability and normal reference values for blood flow parameters. Am J Radiol, 1999, 172:213
  • 6Keats TE, Sistrom C. Atlas of radiologic measurement. A Harcourt Health Science Company 7th ed, 2001, 51
  • 7Lotz F, Meier C, Leppert ,et al. Measurement with phase-contrast MR imaging: basic facts and implementation. Radiolographics, 2002, 22:651
  • 8Ho SSY, Chan YL, Yeung DKW,et al. Blood flow volume quantification of cerebral ischemia: comparison of three noninvasive imaging techniques of carotid and vertebral arteries. AJR, 2002, 178:551
  • 9Ho SSY, Metreweli C. Preferred technique for blood flow volume measurement in cerebrovascular disease. Stroke, 2000, 31:1342
  • 10Mehta NR, Jones L, Kraut MA,et al. Physiologic variations in dural venous sinus flow on phase-contrast MR imaging. AJR, 2000, 175:221

共引文献6

同被引文献53