期刊文献+

用残余气体分析仪测量和计算分压强 被引量:2

Measuring and Calculating the Partial Pressures with Residual Gas Analyzer
在线阅读 下载PDF
导出
摘要 残余气体分析仪可以得到真空系统残余气体的质谱图,并给出各个质荷比峰的离子流值,但不能直接测量残余气体中各组分气体的分压强。在考虑到组分气体对某个峰的贡献可以线性叠加的基础上,本文介绍了峰的离子流与分压强、仪器灵敏度、传输几率和丰度之间的关系,从而得到了利用RGA质谱峰的离子流计算残余气体分压的公式。 Residual Gas Analyzer can get the mass spectrum of residual gases in the vacuum system and give the ion currents of the mass peaks. But it can't measure the partial pressure of each component gas directly. On the grounds of that the contributions of the component gases to a peak can be linearly superposed, this paper introduces the relations between the ion currents of mass peaks and the partial pressures along with the sensitivity of instrument, the transmission factors and the abundances, and then derivates the formulas of calculating partial pressures with ion currents of mass peaks.
作者 肖琼 彭晓华
出处 《真空电子技术》 2008年第5期16-17,21,共3页 Vacuum Electronics
关键词 残余气体分析 离子流 分压强 真空系统 电子倍增器 Residual gas analyzer Ion current Partial pressure
  • 相关文献

参考文献3

二级参考文献5

  • 1康小录.一组四极质谱计的高压力特性[J].真空与低温,1993,12(3):147-151. 被引量:4
  • 2麦克拉弗蒂著 王光辉 姜龙飞 王聪慧译.质谱解析[M].北京:化学工业出版社,1980..
  • 3张浩.[D].北京大学,2001.
  • 4查良镇.质谱学导论讲义[Z].清华大学无线电系,1982..
  • 5张兆祥.质谱数据分析[J].真空科学与技术,1984,4(5):361-361.