本传感器采用Clark电极,传感器性能稳定,重复性好,氧分压在0~120mmHg柱范围内呈良好线性关系,连续测定4小时,其飘移信号小于2mmHg柱。可连续监测病员氧分压的变化。
A sensor using a Clark electrode is described which can monitor the changes of transcutaneous oxygen of patients continuously. The stability and reproducibility are satisfactory. The sensor gives good linear response for oxygen partial pressure in the range of 0-120mm Hg. The drift of the signal is less than 2mm Hg after 4 hours of successive determinations.