期刊文献+

免试剂离子色谱法同时测定饮用水中9种阴离子 被引量:6

Determination of nine anions in drinking water by reagent-free ion chromatography.
在线阅读 下载PDF
导出
摘要 建立了免试剂离子色谱法同时测定饮用水中F^-、BrO_3^-、Cl^-、NO_2^-、SO_4^(2-)、Br^-、NO_3^-、ClO_3^-、PO_4^(3-)等9种无机阴离子的方法。所用仪器为戴安ICS-90离子色谱仪,色谱柱为IonPac AS11-HC,淋洗液氢氧根的浓度为29mmol/L,流速为1mL/min。在所选实验条件下,分析方法线性相关系数的平方为0.9992~0.9998,相对标准偏差为1.20%~4.39%,加标回收率为96.1%~103.8%,最低检出限为0.0012~0.1710mg/L。方法具有背景电导低、灵敏度高、测定结果准确等特点,适合水样中多种阴离子的同时测定。 Nine anions (F^- , BrO3^- , Cl^- , NO2^- , SO4^2- , Br^- , NO3^- , ClO3^- and PO4^3- ) in drinking water were determined using reagent-free ion chromatography. The instrument used was a Dionex ICS-90 ion chromatograph. IonPac ASII-HC was used as spearation column with 29 mmol/L KOH as eluent. The flow rate was 1mL/min. Under the selected experimental conditions, good correlationship was obtained with R^2 = 0. 9992 - 0. 9998 . TheRSDwerel. 20%-4.39% and the recoveries of standard addition were 96.09%- 103.75%. The method has the advantages of low conductivity background, high senstivity and high accuracy and is suitable to determine multiple inorganic anions in water simultaneously.
出处 《分析仪器》 CAS 2008年第1期33-36,共4页 Analytical Instrumentation
  • 相关文献

参考文献7

二级参考文献23

  • 1刘丽菁,蔡一新,倪蕾.淋洗液自动发生-离子色谱法测定水中的氟、氯、硫酸盐和硝酸盐氮[J].中国卫生检验杂志,2004,14(5):576-577. 被引量:5
  • 2国家环境保护总局 水和水和废水监测分析方法编委会 编.水和废水监测分析方法[M](第4版)[M].北京:中国环境科学出版社,2002.243-244.
  • 3Clifford D,Subramonian S.Removing dissolved inorganic contaminants from water[J].Environ Sci Technol,1986,(11):1072-1078.
  • 4WHO. Guidelines for drinking-water quality. Volume 2 Health criteria and other supporting information, 1996. 822-828
  • 5Rook JJ. Formation of haloforms during chlorination of natural waters. Water Treat Exam, 1974, 23(2): 234-243
  • 6王建平.国家计量检定规程汇编(化学)[M].北京:中国计量出版社,1990.300-301.
  • 7武汉大学.分析化学[M].北京:高等教育出版社,1995.391.
  • 8Small H.Twenty years of Ion chromatography[J].International ion chromatography symposium,1990,504:211-245.
  • 9Greengerg A E,Clesceri L S,Eaton A D.Standard methods for the examination of water and wastewater.18th ed[M].Washington D C:Public Health Assoc,1992:128-129.
  • 10D4327-97,Standard Test Method for Anions in water by Chemically Suppressed Ion Chromatography[S].