调强放疗中小野剂量学特性研究 被引量:7

Study on the dosimetric characteristics of small segment dimensions in intensity modulated radiation therapy
原文传递
导出
摘要 目的对加速器小野的剂量学特性进行研究,以指导调强放疗在临床上的应用。方法计算西门子primusM加速器小野条件下的总散射因子并比较相对剂量分布的测量和计算结果。结果在小野测试中,当治疗计划系统计算精度优于0.2 cm时,>2 cm×2 cm射野的总散射因子的计算值误差<3%,所有射野的计算与测量的相对剂量分布均较为吻合,但<2 cm×2 cm的射野在边缘低剂量区有一定的偏差。结论对于治疗计划系统计算的小射野剂量分布的误差在调强计划设计上应加以考虑,多叶准直器的到位精度对小射野剂量影响不容忽视。 Objective To investigate the dosimetric characteristics of small segment dimensions, so as to instruct the clinical application in intensity modulated radiation therapy(IMRT). Methods Using the Siemens primusM accelerator, we compare the results of measurement and calculation about total scatter factors and relative dose distributions in small segments. Results In the small segment dimensions testing, when the calculation precision of TPS was under 0.2 cm, deviations of calculated total scatter factors was within 3% for fields collimated up to 2 cm × 2 cm, calculated relative dose distributions in all fields agree well with measured ones, but there were some deviations for the fields down to 2 cm×2 cm in low dose region. Conclusion The error of TPS calculated dose distributions for small fields must be taken into account, in addition, MLC leaf positional accuracy is important for the small fields, so QA/QC for static MLCs can't be ignored.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2007年第6期469-472,共4页 Chinese Journal of Radiation Oncology
关键词 调强放射疗法 小野 剂量学 Intensity modulated radiation therapy Small segment dimensions Dosimetry
  • 相关文献

参考文献8

  • 1Ezzell GA, Galvin JM, Low DA, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee. Med Phys, 2003, 30: 2089-2115.
  • 2Kung JH, Chen GT. Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy. Med Phys, 2000, 27: 1617-1622.
  • 3Sharpe MB, Miller MB, Yan D, et al. Monitor unit settings for intensity modulated beams delivered using a step-and-shoot approach. Med Phys, 2000, 27: 2719-2725.
  • 4Chui CS, Spirou S, and LoSasso T. Testing of dynamic multileaf collimation. Med Phys, 1996, 23 : 635-641.
  • 5Low DA, Sohn JW, Klein EE, et al. Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. Med Phys, 2001, 28: 752-756.
  • 6马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 7Childress NL,and Rosen Ⅱ. The design and testing of novel clinical parameters for dose comparison. Int J Radiat Oncol Biol Phys, 2003, 56 : 1464-1479.
  • 8Low DA, Harms WB, Mutic S, et al. A technique for the quantitative evaluation of dose distributions. Med Phys, 1998, 25: 656-661.

二级参考文献4

  • 1Lossasso T, Chui CS, Ling CC. Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. Med Phys, 2001, 28:2209-2219.
  • 2Essers M, de Langen M, Dirkx MLP, et al. Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation. Radiother Oncol, 2001,60:215-224.
  • 3Budgell GJ, Mort JH, Williams PC, et al. Requirements for leaf position accuracy for dynamic multileaf collimation. Phys Med Biol, 2000, 45:1211-1227.
  • 4Saw CB, Ayyangar KM, Thompson RB, et al. Commissioning and quality assurace for MLC-based IMRT. Med Dosim, 2001, 26:125-133.

共引文献9

同被引文献29

引证文献7

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部