计算机X线摄影图像的静止滤线栅伪影分析 被引量:2

Analysis of stationary grid artifact on computed radiographic image
原文传递
导出
摘要 目的通过分析计算机 X 线摄影(CR)系统的采样频率和栅密度频率,提出尽可能有效避免滤线栅伪影出现的抑制方法。方法实际测试不同栅密度的滤线栅与不同规格的成像板(IP)匹配使用,得到不同效果的图像;模拟2种信号频率(即2种栅密度的滤线栅),采用3种采样频率[6、8、10图像空间分辨率(pixels/mm)],得到不同的模拟图像效果。结果通过对比分析模拟图和实际图像,发现在采样频率正好等于信号频率的2倍时,可以得到正确的信号频率,图像清晰,无混淆现象。栅密度<4 LP/mm 时,可较好地与14 in×14 in、10 in×12 in(1 in=2.54 cm)IP 匹配使用。结论 CR 系统与普通 IP 匹配使用合适的采样频率和栅密度频率,可以有效避免滤线栅伪影出现。 Objective To analyze sampling frequency and grid density frequency on the computed radiography (CR) system and to develop an effective method to suppress or eliminate grid artifacts. Methods To test grids with different grid-density in combination with IP boards of different specifications. Radiographic images with various qualities were obtained by simulating two kinds of signaling frequencies (namely two kind of grids with different grid densities), and utilizing three kinds of sampling frequencies (6, 8, 10 pixels/mm). A variety of different simulation images were obtained. Results By comparing simulation images with actual images, it was discovered that correct signaling frequency could be achieved if the sampling frequency were equal to twice the signaling frequency. The obtained image was clear and free of grid artifact. A grid density of 〈4 LP/mm was most suitable for IP boards of 14 in × 14 in, 10 in × 12 in, in size. Conclusion Appropriate combination of the CR System and IP boards with matching sampling frequency and grid-density frequency can effectively eliminate or reduce grid artifacts.
作者 王林 滕皋军
出处 《中华放射学杂志》 CAS CSCD 北大核心 2007年第11期1249-1251,共3页 Chinese Journal of Radiology
关键词 放射摄影术 设备失效 投射技术 光密度测定法 X线 Radiography Equipment failure Projective techniques Densitometry,X-ray
  • 相关文献

参考文献9

二级参考文献9

  • 1曾勇明,陈亮,陆云峰.CR图像噪声的原因及抑制方法[J].中国医学影像技术,2004,20(Z2):148-149.
  • 2李泽洪,李祥林,于俊风,王爱玲,李建平.CR系统的科学使用与维护保养探讨[J].医学影像学杂志,2005,15(2):174-175. 被引量:8
  • 3孟代英.X线投照技术[M].第2版.济南:山东科学技术出版社,1992.475-478.
  • 4Williams. 图形程序开发人员指南[M]. 北京:电子工业出版社,1995.
  • 5Acceptance Test and Quality Control of Photo Stimulable Phosphor Imaging Systems,Report task Group #10,AAPM October 1997.
  • 6Protocol for the QA of Computed Radiography Systems,Commissing and Annual QA Tests,KCARE(King's Centre for Assessment of Radiological Equipment)CR Protocol,2003.
  • 7Samei E,Seibert JA,Willis CE,et al,Performance evaluation of computed Radiology systems[J].Med Phys,2001,28:361-371.
  • 8唐荣锡等.计算机图形学教程[M]科学出版社,2000.
  • 9王书敏,宋少娟.CR摄影影像质量的保证[J].医学影像学杂志,2003,13(9):660-660. 被引量:9

共引文献12

同被引文献9

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部