摘要 提出了一种通过无线传输实现血氧饱和度测量的无线血氧指夹。其无线功能主要依靠射频芯片nRF905,并利用TI公司的16位单片机MSP430F1232作为控制芯片,用于配置nRF905以完成信号发射、接收的功能,并通过其内部集成的10位ADC采集经放大、滤波后的血容积率信号。这种无线血氧指夹体积小,避免了长导线的麻烦,适用于无线监护系统。结果表明:原理样机经测试通信距离远,数据传输稳定,所传输数据满足计算血氧饱和度要求。 A design for wireless pulse oximetry finger-tip is introduced. The function of wireless is based on nRF905, The 16 bit MCU, MSP430 are used to control the system, The MCU configures nRF905 to transmit and receive data. The signal of pulse oximetry out of the amplifier and filter is sampled by 10 bit ADC on chip, The wireless pulse oximetry finger-tip can be developed to wireless monitoring system. The result shows that the prototype has been tested. The communication distance is long enough and the data transmitted is steady and meets the requirement of calculating of pulse oximetry.
出处 《生物医学工程研究》 2007年第2期204-207,共4页 Journal Of Biomedical Engineering Research
关键词 无线 血氧饱和度 NRF905 MSP430 监护 Wireless Pulse oximetry nRF905 MSP430 Monitoring
  • 相关文献

参考文献4

二级参考文献12

  • 1孙卫新,金捷,汪家旺,杨新平,陈健.脉搏血氧仪的研制和定标方法探讨[J].西安医科大学学报,1995,16(2):218-221. 被引量:9
  • 2姜宗义,龚卫宁.脉搏血氧饱合度测量[J].医疗卫生装备,1995,16(3):1-4. 被引量:8
  • 3石梅生 赫宪令.无创脉搏血氧仪的研制[J].医疗卫生装备,1995,6:1-3.
  • 4[2]SHIGA T,TANABE K,NAKASE Y,et al.Development of portable tissue oximeter using near infra-red spectroscopy[J].Med Biol Eng Comp,1995,33:622-626.
  • 5[1]YAMAKOSHI K,SHIMAZU H,SHIBATA M,et al.New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger[J].Med Biol Eng Comp,1982,20:307 -313.
  • 6孙卫新,国外医学.生物医学工程分册,1997年,2期,69页
  • 7Strieoei HW, Kretz FJ. The functional principle, reliability and limitations of pulse oximetry[J]. Anaesthesia, 1989,38(12):649-657
  • 8Baker SJ, Tremper KK. The effects of motion on the performance of pulse oximeters in volunteers[J]. Anesthesiology,1997,86(1):101-108
  • 9de Kock JP, Tarassenko L. Pulse oximetry:Theoretical and experimental model[J]. Med Biol Eng Comput,1993,31(3):291-298
  • 10Mendelson Y. Pulse Oximetry:Theory and application for noninvasive monitoring[J]. Clin Chem,1992,38(9):1601-1607

共引文献48

同被引文献24

引证文献4

二级引证文献38

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部