期刊文献+

预稀释标本放置时间对血细胞分析仪测定结果的影响 被引量:14

Efects of pre diluted blood specimen lay up time on hematology analyzer measuring results
暂未订购
导出
摘要 分析了血液标本预稀释后放置0,20,30,60和180分钟血细胞分析仪测定结果的变化情况。发现RBC、WBC、HGB和MPV四项参数的测定结果基本不发生改变。但是WBC三分类的结果和WBC分布曲线的形态随稀释标本放置时间延长会发生明显改变,在30min时,LYM%明显升高(P<0.05),GRAN%明显降低(P<0.05);PLT和MCV随稀释标本放置时间延长而逐渐降低,PLT降低在60min,MCV降低在30min开始有统计学意义(P<0.05)。因此,使用血细胞分析仪时,已稀释标本应在20min以内测定。 The test results by hematology analyzer at 0,20,30,60 and 180 minutes after the dilution of anti coagulating blood specimen were studied in this paper.It was observed that RBC,WBC,Hb and MPV results of pre diluted specimen were stable in 180 minutes,but MCV,PLT and WBC 3 parts differential values as well as WBC histogram changed with the diluted time of the samples.The writer suggest that pre diluted blood specimen be tested in 20 minutes.If any pre diluted sample is tested later than 20 minutes,the WBC 3 parts differential and PLT results can not be reported.
作者 吕斌斌
出处 《临床检验杂志》 CAS CSCD 北大核心 1997年第1期32-34,共3页 Chinese Journal of Clinical Laboratory Science
关键词 预稀释血标本 血细胞分析仪 放置时间 Pre diluted blood specimen Hematology analyzer Lay up time.